System Cleaning Instructions 

The generator and converter may be cleaned using an acid-free cleaning solution (i.e. glass 
cleaner).   
 
Probes should be cleaned using isopropyl alcohol.  Probes are made from titanium and can be 
autoclaved (the converter is an electrical part and cannot be sterilized in this manner).   Before 
each procedure place the probe tip in water or alcohol and turn the power on for a few seconds 
to remove residue.  The tip also can be sterilized using alcohol with the power on.   

 
 
 

9.  Troubleshooting 
 

Your Ultrasonic Processor was designed to provide you with years of safe and dependable 
service. Nevertheless, because of component failure or improper usage, the possibility does 
exist that it might not perform as it should, shut down or stop working all together. The most 
probable causes for malfunction are listed below and should be investigated. 
 

  A connector or cable is damaged.  

 

  The unit was plugged into an electrical outlet that provides a different voltage from that 

required. See 

Electrical Requirements.

 

  The microtip is not tightened properly with the wrenches provided.

 

  The convertor and/or microtip has been dropped.

 

  A microtip being operated is damaged or worn past its useful life.

 

  A fuse(s) has failed. If a fuse(s) has failed, proceed as follows:

 

1.  Turn the unit off by depressing the OFF key, and disconnect the line cord from the 

electrical outlet. 

2.  Open the fuse holder cover using a small screwdriver, and pull out the red fuse 

holder from its housing. 

3.  Replace the fuse(s). 
4.  Reconnect the line cord to the electrical outlet, press the ON key and set the 

amplitude to 100. With the probe in air (out of the sample), the wattmeter should read 
below 10 watts. If the reading exceeds 10 watts, press the OFF key, and disconnect 
the probe from the converter. 

5.  Press the ON key. If the wattmeter reads below 10 watts, the probe has failed or is 

out of tune due to excessive erosion, and should be replaced, if the wattmeter reads 
above 20 watts, either the converter or power supply has failed and the complete 
Ultrasonic Processor should be returned for repair. 

 

 
OVERLOAD CONDITION 

If the Ultrasonic Processor stops working, and an OVERLOAD indication is displayed on the 
screen, check for possible causes as outlined in the above paragraph, then press the 

OFF

 key 

to switch the unit off, and the 

ON

 key to switch the unit back on. 

 

 

If the problem persists after inspecting all of these, please contact Customer Service for 
additional assistance or to replace a worn microtip or damaged part. 

 

Содержание Sonicator

Страница 1: ...11 Sonicator ULTRASONIC PROCESSOR Part No Q125 OPERATION MANUAL 53 Church Hill Road Newtown CT 06470 U S A Phone 203 426 0101 1 877 338 9636 Fax 203 426 7026 Web www sonicator com E mail info sonicato...

Страница 2: ...Rev 3 11 2...

Страница 3: ...pecifications 6 7 4 Principles of Operation 8 9 5 Description of Components Functions of Controls 10 12 6 Preparation for Use 13 7 Operating Instructions 14 17 8 Maintenance 18 20 9 Troubleshooting 20...

Страница 4: ...rranty of either merchantability or fitness for a particular purpose No person or company is authorized to change modify or amend the terms of this warranty in any manner or fashion whatsoever Under n...

Страница 5: ...Generator within a Fume Hood Hearing protection is highly recommended It is recommended that a sound abating enclosure or ear protection be used when operating the Ultrasonic Processor NEVER immerse...

Страница 6: ...slo blo 1 6 Amps slo blo Weight 7 lbs 3 2 Kg Dimensions 8 W x13 75 L x 5 75 H 203mm x 349mm x 146mm Output Voltage 1000 V rms max Output Frequency 20 KHz Converter Weight 0 75 lbs 0 34Kg Dimensions 6...

Страница 7: ...e used If the 220V plug is not configured to match the wall receptacle a properly grounded universal AC socket adapter must be added Important Universal adapters do not convert voltage or frequency Fi...

Страница 8: ...ore information on this subject Please consult with a product specialist for assistance with selecting a probe for your application Relationship of Amplitude and Wattage Sonication power is measured i...

Страница 9: ...rvation the amount of power required cannot be predetermined A sensing network continuously monitors the output requirements and automatically adjusts the power to maintain the amplitude at the presel...

Страница 10: ...f Controls The Model Q125 includes a standard 1 8 diameter microtip 4422 Q125 Front Panel Q125 Rear Panel Generator Converter 1 8 Microtip Converter Cable Converter Cable Connector Power Cord Connecto...

Страница 11: ...main power on 0 key Switches the main power off AMPL Controls the amplitude of vibration at the probe tip key Used with the AMPL key when the unit is on stand by to set the amplitude of vibration at...

Страница 12: ...to pin 6 9 PIN D SUB CONNECTOR Pin No Description 1 Not connected 2 Not connected 3 Not connected 4 Enables connection to a frequency counter 5 Enables connection to an external power monitor 5 mv 1...

Страница 13: ...sary to convert the unit for different voltage operation proceed as follows 1 Ensure that the power cord is not connected to the electrical outlet 2 Open the fuse holder cover using a small screwdrive...

Страница 14: ...the convertor and probe assembly is not already assembled screw the probe onto the convertor and using the wrenches provided tighten it securely See instruction in Maintenance section for attaching an...

Страница 15: ...ative If the process is initiated by the footswitch the STOP key becomes inoperative 3 To increase or decrease the amplitude in small increments when the ultrasonics is on depress the AMPL to display...

Страница 16: ...et independently from 01 second to 59 seconds During the OFF portion of the cycle the red indicator on the PULSE key will illuminate If the OFF portion of the cycle exceeds three seconds a cautionary...

Страница 17: ...processing time and elapsed processing time e g Set 0 30 00 Time 0 22 10 c Selected pulsing cycle and actual pulsing cycle e g Pulse 01 01 01 00 d Amount of power in watts and accumulated amount of en...

Страница 18: ...nance Ultrasonic processors create high intensity vibration which puts stress on the converter and horn assembly The sides and end of the probe must never be allowed to come in contact with anything b...

Страница 19: ...cotton swab to remove any debris on the threading of the connecting stud Allow the alcohol to dry completely 3 Clean threading in converter Use alcohol and a cotton swab to remove any debris on the th...

Страница 20: ...otip being operated is damaged or worn past its useful life A fuse s has failed If a fuse s has failed proceed as follows 1 Turn the unit off by depressing the OFF key and disconnect the line cord fro...

Страница 21: ...ion Number prior to returning the instrument Care should be exercised to provide adequate packing to insure against possible damage in shipment The Ultrasonic Processor should be sent to the Service D...

Страница 22: ...h or exposed to radiological biological or chemical agents and has been decontaminated rendering it safer for handling use or disposal Authorization By accepting authorization to return the equipment...

Отзывы: