background image

Rev. 3-11 

 

18 

8.  Maintenance 
 

It is recommended to periodically inspect the unit, both visually and physically, to insure optimum 
and safe performance.  This inspection should be scheduled as a routine maintenance 
procedure, done with the unit power OFF and with the unit unplugged from the AC power source. 
 
Long exposure to acids or caustics results in corrosion of metal parts or components. Check the 
generator, converter, and cables periodically for any signs of rust or discoloration.  If 
discoloration is found, move the ultrasonic processor away from the source of the contaminant. 
 
Examine the condition of the high voltage cable that attaches the converter to the generator.  
Inspect the wire insulation for damage, such as wear, burning from hot plate contact or breakage 
from extended use or rough handling.  In general use, the cable assembly should not be used to 
carry the converter or pull it toward the user.  Make certain the cable always has slack and is 
never tensioned.  If necessary, move the generator or converter assembly closer to one another 
to accomplish this.  

WARNING:  Do not use a cable with broken end connections, exposed 

wires or frayed insulation.  High voltage is present in the cable and will pose a shock 
hazard.  Do not touch the converter assembly until the power switch is off and the unit is 
unplugged.  
 
Microtip/ Probe Maintenance 

Ultrasonic processors create high intensity vibration which puts stress on the converter and horn 
assembly.   The sides and end of the probe must 

never

 be allowed to come in contact with 

anything but the solution.  When using a Microtip, the stress resulting at the point of contact with 
the vessel could cause the Microtip to fracture.   
 
Proper care of the probe is essential for dependable operation.  The intense cavitation will, after 
usage for period of time, cause the tip to erode, and the power output to decrease.  The 
smoother and shinier the tip, the more power will be transmitted into the sample. The vibrations 
may also cause the probe tip to loosen over time or the threaded connection to accumulate 
debris.  

Note:

  

A loose probe will usually generate a loud piercing or squealing sound.   

 

For that reason, 

it is recommended that a preventative maintenance schedule be adopted 

to examine the unit at regular intervals.

  The schedule should depend on frequency of use.  

Weekly maintenance schedules are recommended for units used frequently or monthly for those 
used infrequently.  The tip must be examined for excessive wear and to ensure that the 
threaded connection is clean and attached properly to the convertor.  Use a cotton swab and 
alcohol (i.e. ethanol, isopropyl, etc.) to clean the threaded mating surfaces.

    

 
 
When excessive wear (corrosion/pitting of the probe tip) is detected the probe should be 
replaced with a new one.

 

 
 

WARNING:  Never hand tighten probes or horns onto the convertor; properly tighten 
them with the appropriate Wrench Set.   
 

 
 
 
 

Содержание Sonicator

Страница 1: ...11 Sonicator ULTRASONIC PROCESSOR Part No Q125 OPERATION MANUAL 53 Church Hill Road Newtown CT 06470 U S A Phone 203 426 0101 1 877 338 9636 Fax 203 426 7026 Web www sonicator com E mail info sonicato...

Страница 2: ...Rev 3 11 2...

Страница 3: ...pecifications 6 7 4 Principles of Operation 8 9 5 Description of Components Functions of Controls 10 12 6 Preparation for Use 13 7 Operating Instructions 14 17 8 Maintenance 18 20 9 Troubleshooting 20...

Страница 4: ...rranty of either merchantability or fitness for a particular purpose No person or company is authorized to change modify or amend the terms of this warranty in any manner or fashion whatsoever Under n...

Страница 5: ...Generator within a Fume Hood Hearing protection is highly recommended It is recommended that a sound abating enclosure or ear protection be used when operating the Ultrasonic Processor NEVER immerse...

Страница 6: ...slo blo 1 6 Amps slo blo Weight 7 lbs 3 2 Kg Dimensions 8 W x13 75 L x 5 75 H 203mm x 349mm x 146mm Output Voltage 1000 V rms max Output Frequency 20 KHz Converter Weight 0 75 lbs 0 34Kg Dimensions 6...

Страница 7: ...e used If the 220V plug is not configured to match the wall receptacle a properly grounded universal AC socket adapter must be added Important Universal adapters do not convert voltage or frequency Fi...

Страница 8: ...ore information on this subject Please consult with a product specialist for assistance with selecting a probe for your application Relationship of Amplitude and Wattage Sonication power is measured i...

Страница 9: ...rvation the amount of power required cannot be predetermined A sensing network continuously monitors the output requirements and automatically adjusts the power to maintain the amplitude at the presel...

Страница 10: ...f Controls The Model Q125 includes a standard 1 8 diameter microtip 4422 Q125 Front Panel Q125 Rear Panel Generator Converter 1 8 Microtip Converter Cable Converter Cable Connector Power Cord Connecto...

Страница 11: ...main power on 0 key Switches the main power off AMPL Controls the amplitude of vibration at the probe tip key Used with the AMPL key when the unit is on stand by to set the amplitude of vibration at...

Страница 12: ...to pin 6 9 PIN D SUB CONNECTOR Pin No Description 1 Not connected 2 Not connected 3 Not connected 4 Enables connection to a frequency counter 5 Enables connection to an external power monitor 5 mv 1...

Страница 13: ...sary to convert the unit for different voltage operation proceed as follows 1 Ensure that the power cord is not connected to the electrical outlet 2 Open the fuse holder cover using a small screwdrive...

Страница 14: ...the convertor and probe assembly is not already assembled screw the probe onto the convertor and using the wrenches provided tighten it securely See instruction in Maintenance section for attaching an...

Страница 15: ...ative If the process is initiated by the footswitch the STOP key becomes inoperative 3 To increase or decrease the amplitude in small increments when the ultrasonics is on depress the AMPL to display...

Страница 16: ...et independently from 01 second to 59 seconds During the OFF portion of the cycle the red indicator on the PULSE key will illuminate If the OFF portion of the cycle exceeds three seconds a cautionary...

Страница 17: ...processing time and elapsed processing time e g Set 0 30 00 Time 0 22 10 c Selected pulsing cycle and actual pulsing cycle e g Pulse 01 01 01 00 d Amount of power in watts and accumulated amount of en...

Страница 18: ...nance Ultrasonic processors create high intensity vibration which puts stress on the converter and horn assembly The sides and end of the probe must never be allowed to come in contact with anything b...

Страница 19: ...cotton swab to remove any debris on the threading of the connecting stud Allow the alcohol to dry completely 3 Clean threading in converter Use alcohol and a cotton swab to remove any debris on the th...

Страница 20: ...otip being operated is damaged or worn past its useful life A fuse s has failed If a fuse s has failed proceed as follows 1 Turn the unit off by depressing the OFF key and disconnect the line cord fro...

Страница 21: ...ion Number prior to returning the instrument Care should be exercised to provide adequate packing to insure against possible damage in shipment The Ultrasonic Processor should be sent to the Service D...

Страница 22: ...h or exposed to radiological biological or chemical agents and has been decontaminated rendering it safer for handling use or disposal Authorization By accepting authorization to return the equipment...

Отзывы: