20
Programming Screen
How To Create a Program
The Sonicator has the ability to save up to 10 programs.
1. Select a program number from the keypad, the program number will appear above the Amplitude
box.
For each of the following steps
: After touching a field, the background will change to yellow
indicating an active field. After entering data, the entry is completed by touching Enter (ENT), touching
the field a second time or by touching a new field.
2. Program a value into the Amplitude field.
3. Program total Process Time (total active sonication time).
4. Program Pulse On time – If no data is entered (00hr:00min:00sec), the unit will run continuously
without pulsing.
5. Program Pulse Off time – If no data is entered, the unit will run continuously without pulsing.
Process Time, Pulse On and Pulse Off times are measured in Hours: Minutes: Seconds format.
6. Activate Temperature Monitoring (if necessary).
7. Program a Shutdown Temperature (if necessary). Enter a temperature value.
During sonication, if the temperature reaches the Shutdown value, the unit will pause sonication.
The unit will remain paused until the temperature drops below the pre-set value. Once the
temperature drops below the value, sonication can be manually resumed.
8. Touch Save to store the program to the memory number selected in step 1.
9. Touch Run to proceed to the following screen.
10. Touch Start to begin the program.
Program
Progress
Total time of active
sonication that has occurred
during the current program.
Rest or Off time is not
included in this value.
Total programmed time of
active sonication. Rest or
Off time is not included in
this value.
Содержание SONICATOR S-4000
Страница 34: ...34 Cleaning horn stud Cleaning horn threading...
Страница 43: ...43...
Страница 53: ...53...