background image

Rev. 4-16a 

19

 

 

into mechanical energy. It does this by exciting the piezoelectric crystals causing them to move in the 
longitudinal direction within the convertor. This change from electrical into mechanical energy causes a 
motion that travels through the horn/probe causing the tip to move up and down. 
 
The distance of one movement up and down is called its amplitude. The amplitude is adjustable. Each 
probe has a maximum amplitude value.  For example, with a 1/8” diameter probe at setting 100%, the 
probe will achieve an amplitude of approximately 180μm. At setting 50% the amplitude is 
approximately 90μm.  Note: this is approximate and not perfectly linear.  Qsonica measures the 
amplitude of each probe at 100% and these values are published in the brochure. 
 
Amplitude and intensity have a direct relationship. If you operate at a low amplitude setting, you will 
deliver low intensity sonication. If you operate at a high amplitude setting, you will have high intensity 
sonication. In order to be able to reproduce results, the amplitude setting, temperature, viscosity and 
volume of the sample are all parameters that need to remain consistent. The amplitude, not the power, 
is most critical when trying to reproduce sonication results. 
 
Power has a variable relationship with amplitude/intensity. For example, sonicating water requires less 
wattage when compared to a viscous sample (such as oil). While sonicating both samples at the same 
amplitude setting the power/wattage will differ because the viscous sample will require more watts in 
order to drive the horn. The viscous sample puts a heavier load on the probe so they system must work 
harder to vibrate up and down at the same amplitude setting.  The oil may draw double the watts when 
operated at the same amplitude as the water sample. 
 
Small fluctuation in the wattage during sonication is normal.  Major swings in wattage (+/- 30 watts) 
may indicate a problem with the sample, setup or the sonicator itself.

 

 

 

 

Содержание Q55

Страница 1: ...Rev 6 19 Sonicator ULTRASONIC PROCESSOR Part No Q55 OPERATION MANUAL Qsonica LLC 53 Church Hill Rd Newtown CT 06470 USA Phone 203 426 0101 Fax 203 426 7026 info sonicator com www sonicator com...

Страница 2: ...rinciples of Operation 7 5 Description of Components Functions of Controls 8 10 6 Preparation for Use 11 7 Operating Instructions 12 13 8 Maintenance 14 16 9 Troubleshooting 16 10 Service Return of Eq...

Страница 3: ...arranty of either merchantability or fitness for a particular purpose No person or company is authorized to change modify or amend the terms of this warranty in any manner or fashion whatsoever Under...

Страница 4: ...me Hood Hearing protection is highly recommended It is recommended that a sound abating enclosure or ear protection be used when operating the Ultrasonic Processor NEVER immerse the converter in liqui...

Страница 5: ...Voltage 24 VDC Rated Power 55 Watts Weight 3 lbs 1 36 Kg Dimensions 8 W x 7 5 L x 5 75 H 20cm x 19cm x 14 6cm Output Voltage 700 Vrms max Output Frequency 20 KHz Converter Weight 0 75 lbs 0 34Kg Dimen...

Страница 6: ...ot configured to match the wall receptacle a properly grounded universal AC socket adapter must be added Important Universal adapters do not convert voltage or frequency Manufacturer is not responsibl...

Страница 7: ...ation of the tip causes cavitation the formation and violent collapse of microscopic bubbles The collapse of thousands of cavitation bubbles releases tremendous energy in the cavitation field The eros...

Страница 8: ...Components Functions of Controls The Model Q55 includes a standard 1 8 diameter microtip 4422 Q55 Front Panel Q55 Rear Panel Generator Converter 1 8 Microtip Converter Cable Footswitch Jack Convertor...

Страница 9: ...robe tip Continuous Thumb Switch Continuous Position Probe tip is energized when the amplitude is set above 0 ThumbSwitch Position Probe tip is energized once the thumbswitch or footswitch is depresse...

Страница 10: ...Rev 4 16a 10 Note Improper clamping can damage the system and void the warranty Correct way to clamp Not correct Note Using a Qsonica sound enclosure or stand will ensure a proper fit...

Страница 11: ...iversal 100V 240V 50 60 Hz AC to DC switch mode power supply capable of delivering 24 volts at 2 5 amperes is supplied with your system If it is necessary to convert the unit for different voltage ope...

Страница 12: ...ssembled screw the probe onto the convertor and using the wrenches provided tighten it securely Check the tightness of an already assembled probe by using the wrench set Please refer to images in the...

Страница 13: ...ch or Footswitch mode 1 Ensure that the amplitude control knob is set to 0 2 Set the switch on the front panel to ThumbSwitch 3 If needed connect the foot switch cable to the designated jack on the ba...

Страница 14: ...ltrasonic processors create high intensity vibration which puts stress on the converter and horn assembly The sides and end of the probe must never be allowed to come in contact with anything but the...

Страница 15: ...cotton swab to remove any debris on the threading of the connecting stud Allow the alcohol to dry completely 3 Clean threading in converter Use alcohol and a cotton swab to remove any debris on the th...

Страница 16: ...renches provided The convertor and or microtip has been dropped A microtip being operated is damaged or worn past its useful life If the problem persists after inspecting all of these please contact C...

Страница 17: ...or exposed to radiological biological or chemical agents and has been decontaminated rendering it safer for handling use or disposal Authorization By accepting authorization to return the equipment l...

Страница 18: ...d the liquid is pushed down and away in all directions If the vessel is too wide it will not mix effectively and some sample will remain untreated at the periphery Twice the volume in a narrow vessel...

Страница 19: ...nsity sonication If you operate at a high amplitude setting you will have high intensity sonication In order to be able to reproduce results the amplitude setting temperature viscosity and volume of t...

Отзывы: