Rev. 4-16a
15
Follow the steps below for attaching and detaching microtip probes:
1. Disconnect probe from convertor. Use the wrench set provided with the system.
2. Clean threaded stud. Use alcohol and a cotton swab to remove any debris on the threading of the
connecting stud. Allow the alcohol to dry completely.
3. Clean threading in converter. Use alcohol and a cotton swab to remove any debris on the
threading. Do not allow liquid to drip into Convertor. Allow the alcohol to dry completely.
4. Reattach probe to convertor. Screw the probe back onto the convertor and tighten with the wrench
set provided.