8 Operating Your Cleaner
Before You Start Cleaning
Step
Action
1
Select your cleaning solution (refer to page 23 for solution
effects on metals).
2
Allowing for the volume of the parts you will be cleaning and
cleaning solution, fill the tank with warm tap water to the
operating level line (one inch from the top).
3
Add cleaning solution to the tank water.
4
Plug the cleaner into a grounded outlet.
5
For maximum efficiency, refer to page 16, “Optimizing Your
Cleaner” before proceeding.
CAUTION
Do not place parts or containers directly on the bottom of the cleaning
tanks; use a tray or wire to suspend items.
Do not allow the solution to drop more than 3/8 inch below the
operating level line with the cleaner on.
Do not ever use alcohol, gasoline or flammable solutions. Doing so
could cause a fire or explosion. Use only water-based solutions.
Do not ever use mineral acids. These could damage the tank.
Failure to comply with these cautions will void your warranty.
NOTE:
If this is the first time you are running the cleaner, or if you have changed
cleaning solution, you must degas the solution. If not, skip to
“Cleaning
Items”
.