39 MAINTENANCE
'
- - - - -
- - - - -
' Appears in the Display.
This is a normal message indicating the “Leads Off” condition. It will
appear if any electrodes fail to make good contact with the skin. Check
the electrode skin contact until the ECG monitoring mode is established.
If using a multi lead cable, also check any electrode adapters by
swapping them. If a leads off indication occurs consistently for a
particular patient lead after changing the electrode and the adapter, it
may indicate a fault in the lead, in which case please contact your
QRS Diagnostic, LLC service center.
NOTE:
For better direct thorax contact, moisten the back electrodes.
ECG signal appears noisy or shows interference.
Check that the ECG electrodes are well adhered and applied according
to the manufacturer's instructions. Distance the BIOLOG 3000
i
and the
patient from other electrical equipment.
“This cable is incompatible with the 3000i” Message When
Transferring ECG to PC
This can occur if you are touching the back electrodes or are near a
source of electrical noise while the download cable is connected. It does
not interfere with the download operation, however to clear the message
ensure that the Biolog is resting on an insulated surface and away from
power cables, transformers or other electrically noisy equipment.
Unable to take a multi lead ECG with the Biolog Patient cable
Make sure that the cable is connected properly. Examine the cable’s
plug for bent pins. Ensure that the Biolog is configured for simultaneous
cable support on the setup screen.
Read the instructions for multi lead ECGs carefully. If you press the
“Hold” button, only a single lead will be stored. To acquire a multi lead
ECG, the “Mode” button must be pressed and held until acquisition mode
is entered.
Biolog emits a constant sound
The Biolog will emit a faint sound when the LCD backlight is on. If the
backlight is not needed, switch it off using the setup screen (this will
greatly extend the battery life).