QIAxcel Connect Safety and Installation Guide 10/2021
43
Preparing markers
1.
Load 15 μl QX (HS) Alignment Marker into each tube of a QX 0.2 ml 12-Tube Strip.
Add 1 drop of mineral oil to each tube and place the strip into the
MARKER1
position of
the buffer tray.
2.
Load 15 μl QX (HS) Intensity Calibration Marker into each tube of a QX Colored 0.2 ml
12-Tube Strip. Make sure there are no air bubbles, add a drop of mineral oil to each tube,
and insert the strip into the
MARKER2
position of the buffer tray.