8 Camera Maintenance
Drying the CCD Chamber
The CCD sensor is located in a CCD chamber. This chamber is airproof and connect to the socket
of the silicone gel tube (drying tube) on one side of the case. If there is moisture and it causes
the CCD sensor glass to get dew. You can use the silicone gel tube (drying tube) included in the
package to dry it.
Please put the effective silicone gel into the silicone gel tube. Make sure to put a small amount of
cotton inside to prevent the silicone gel get into the CCD chamber.
The process of drying usually needs 24 hours.
Avoid CCD chamber optic window get dew
If the environment humidity is very high, the optic window of the CCD chamber may get dew.
The QHY16200A has the build-in heat board on this optic window to heat the optic window to
avoid it. In most conditions it can take effect.
If the optic window still get dew please try the following method.
1. Avoid the CCD Camera face down to ground. The cold air is heavy than warm air, if you put the
camera facing down to the ground, the cold air is easy to get the optic window and cause it very
cold and get dew
2. Increase the temperature of the CCD sensor. You can increase the CCD sensor temperature a
little to prevent the optic window get dew.
3. Check if the heat board working. If the heat board not working, the optic window is easy to get
dew. Normally the heat board can get about 65-70C in +25C environment. If it is not so warm, it
maybe the problem of the heat board damaged. Need to contact QHYCCD to replace one heat
board.
Содержание QHY16200A
Страница 1: ...All in one COLDMOS Astronomy Camera QHY16200A User s Manual Rev 1 0...
Страница 11: ...QHY16200A...
Страница 14: ...QHY16200A In the window shown below you can set up your filter wheel...
Страница 15: ...QHY16200A Keep clicking OK to return to the Camera Control window Click Connect...
Страница 16: ...QHY16200A...
Страница 20: ...QHY16200A...
Страница 26: ...QHY16200A You need to prepare the tools by yourself to open the camera...
Страница 27: ...QHY16200A Remove all the screws on front of the camera case and remove the front case...
Страница 28: ...QHY16200A You will see two screws to fix the filter wheel case the middle layer of the camera Unscrew them...
Страница 29: ...QHY16200A You will see the power PCB board Remove the two screw of the power PCB board...
Страница 30: ...QHY16200A Remove the three cable carefully Please use a film to protect the CCD chamber optic window...
Страница 33: ...QHY16200A Now the CCD chamber is removed The CCD sensor can be clean here...
Страница 40: ...QHY16200A...
Страница 41: ...QHY16200A When connected with the OAG M The back focus is 33 5 34 13 46 5mm 47mm...
Страница 42: ...QHY16200A 11 Revision history Date of change Revision Content of change January 11 2017 1 0 First edition...