4. What's small display on the camera for?
The OLED display is used for display certain formation related to the shooting. It is
directly driven by application program. So it may display different information or
doesn't display anything at all based on different application programs. Currently all
application program don't support this display, please wait for latest application
program that supports this display.
5. What is defect columns?
You may see one or more vertical columns whose top pixel is a bright dot. Normally
speaking such columns are not defect column. It is resulted from one pixel with high
thermal current. All the pixels below that pixel will be affected by it as thermal
current is transferred through that pixel.
This pixel can be removed by dark frame calibration. If they cannot be eliminated
entirely, you can solve this problem by using some astronomical image processing
software. These software will replace that column with the average value of its
surrounding columns.
Aside from the columns we just talked about, you may see one or more black
columns which is not full sensitive to the light or it is dim to light. This is the defect
column and it cannot be removed by dark frame calibration. The KAF16200 sensor
has two grade, grade1 and grade2. From the specification of the datasheet, the
grade 2 allows some defect columns. For more information about this please check
KAF16200 datasheet. You can use the "defect tools" in some image process software
to remove such defect columns as well.
The following image shows a typical "defect" columns caused by a hot pixel, which is
not very bright and it is usually not considered as a defect column. It will be gone
after dark frame calibration.
Содержание QHY16200A
Страница 1: ...All in one COLDMOS Astronomy Camera QHY16200A User s Manual Rev 1 0...
Страница 11: ...QHY16200A...
Страница 14: ...QHY16200A In the window shown below you can set up your filter wheel...
Страница 15: ...QHY16200A Keep clicking OK to return to the Camera Control window Click Connect...
Страница 16: ...QHY16200A...
Страница 20: ...QHY16200A...
Страница 26: ...QHY16200A You need to prepare the tools by yourself to open the camera...
Страница 27: ...QHY16200A Remove all the screws on front of the camera case and remove the front case...
Страница 28: ...QHY16200A You will see two screws to fix the filter wheel case the middle layer of the camera Unscrew them...
Страница 29: ...QHY16200A You will see the power PCB board Remove the two screw of the power PCB board...
Страница 30: ...QHY16200A Remove the three cable carefully Please use a film to protect the CCD chamber optic window...
Страница 33: ...QHY16200A Now the CCD chamber is removed The CCD sensor can be clean here...
Страница 40: ...QHY16200A...
Страница 41: ...QHY16200A When connected with the OAG M The back focus is 33 5 34 13 46 5mm 47mm...
Страница 42: ...QHY16200A 11 Revision history Date of change Revision Content of change January 11 2017 1 0 First edition...