Troubleshooting
If the temperature of the cabinet regularly, or persistently reaches outside of the official Controlled
Room Temperature Range (+15°C – +25°C), check the environment in which it is installed. Extremes
of ambient temperature may affect the rate of change of the internal temperature and extremes
may lead to high or low alarm states.
If the environment is suitable or hasn’t changed since before the problem, make a note of the details
(chamber temperature, high and low temperatures, ambient temperature, any alarms etc) and call
your supplier for technical assistance.
If the cooling system becomes damaged and the cabinet fails to cool, disconnect the appliance from
the power supply and call your supplier for technical assistance. All work on, or repairs to the cooling
system must be performed by a competent, fully qualified refrigeration engineer.
Error Codes
:
E1 or – E1
Probe error, call for technical assistance
EPr
Internal memory error, press P
MCRTL21.1 7