The controller is preset and in most cases, it is not necessary to adjust the settings.
Normally, the screen will show the current temperature within the cabinet.

High / Low Temperature Monitoring

The controller records the highest and lowest temperatures monitored.

To view the stored values, press 

.

The display will flash 

followed by the lowest recorded temperature.

Press 

again to display the high value.

The display will flash 

followed by the highest recorded temperature.

If 

is pressed again, the display will flash 

followed by the current chamber temperature.

Pressing 

again will return to 

and the menu will cycle as described above.

If the buttons are not touched for

10 seconds, the screen will return to normal

Resetting the Recorded High / Low Values

Display the value that you would like to reset ( 

or 

, as described above). While it is showing,

press and hold 

until the screen shows 

.

Please note that each value ( 

and 

) is reset individually.

Control

The digital controller is placed in the control panel, see fig 2

fig 2

4                                                                                                                                                                          MCRTL21.1

Содержание CRT410

Страница 1: ...USER MANUAL CRT410 CRT640 CRT900 CRT1250 CONTROLLED ROOM TEMPERATURE CABINETS ...

Страница 2: ...ces or anæsthetic or other flammable gasses Power supply The cabinet is designed for 220 240 V 50 Hz The appliance must be correctly earthed The cabinet is fitted with a B S 1363 UK mains plug A 13 amp fuse must be used The flexible cord fitted has three cores The colours used are Green Yellow Earth Blue Neutral Brown Live Do not use this appliance with multi socket distribution extension leads Ex...

Страница 3: ... that reaches extremes of cold for example in goods bay where the temperature is largely affected by the outside may affect the internal temperature regulation Important Before it can be turned on for the first time the transport safety strap must be removed from the compressor Access the compressor compartment from behind the unit The strap is held on by two cross headed screws at either end see ...

Страница 4: ...ay will flash followed by the highest recorded temperature If is pressed again the display will flash followed by the current chamber temperature Pressing again will return to and the menu will cycle as described above If the buttons are not touched for 10 seconds the screen will return to normal Resetting the Recorded High Low Values Display the value that you would like to reset or as described ...

Страница 5: ... desired from the main screen press P The screen will alternate between and the current set point Use the arrows to adjust it and press P again to confirm Please note that the set point is limited to between 17 and 21 C If the monitored temperature is regularly reaching outside Controlled Room Temperature limits causing high or low alarms please call your supplier for technical assistance Please n...

Страница 6: ... cleaning If it becomes damaged and the cabinet fails to cool please call your supplier for technical assistance Disconnect from the mains power supply before cleaning Monthly use only a light soap solution to clean the internal and external surfaces of the cabinet Do not use any abrasives or cleaning product containing chlorine or other harsh chemicals as they can damage the surfaces and the cool...

Страница 7: ...t changed since before the problem make a note of the details chamber temperature high and low temperatures ambient temperature any alarms etc and call your supplier for technical assistance If the cooling system becomes damaged and the cabinet fails to cool disconnect the appliance from the power supply and call your supplier for technical assistance All work on or repairs to the cooling system m...

Страница 8: ...Unit 21 Botany Business Park Macclesfield Road Whaley Bridge High Peak SK23 7DQ T 01663 735494 www qedscientific com MCRTL21 1 ...

Отзывы: