Alarms

The controller is preset with an alarm that will sound, should the monitored temperature go
outside of the official Controlled Room Temperature range (+15°C – +25°C) for more than 2
minutes.

*

In the case of an alarm, the screen will show 

or 

(according to the nature of the alarm) and the

alarm LED (  ) will light.
To silence the sounder, press 

. If the alarm condition continues for some time, the sounder will

resume (it can be silenced again if required)

There are  voltage free contacts on the rear of the unit, which can be connected to an external
alarm system if required. These will activate whenever the unit is in an alarm state. There are
three contacts: COM, NO and NC. To attach these to an external system, contact your alarms
systems administrator.

Set Temperature

The cabinet is preset to control with the official Controlled Room Temperature range (+15°C –
+25°C). As control is based on measured temperature, rather than heating / cooling power, in
normal circumstances it will not be necessary to change the set point (control temperature) of the
controller.

To change the set point, if desired, from the main screen press P. The screen will alternate between

and the current set point. Use the arrows to adjust it and press P again to confirm. Please note
that the set point is limited to between 17 and 21 °C

If the monitored temperature is regularly reaching outside Controlled Room Temperature limits
(causing high or low alarms), please call your supplier for technical assistance.

Please note that the ambient temperature around the cabinet will affect the rate of change of the
internal temperature and extremes may lead to high or low alarm states.

Regular calibration and servicing is recommended to ensure continued correct operation and so as
to avoid reducing the life of the cabinet.

Defrosting

Due to the temperature range of the cabinet, regular frosting of the evaporator plate should not be
an issue, however it is possible to perform a short off-cycle defrost. Should you need to force a
defrost, press and hold 

until the defrost LED (  ) lights.

Lock

For security, the cabinet is fitted with a key lock (in the control panel). Please keep the key(s) safe
and keep a record of any key numbers so that replacements can be issued if required.

*

To allow the cabinet to reach temperature, the alarm is inhibited for the first 45 minutes after start up.

MCRTL21.1                                                                                                                                                                         5

SP

Содержание CRT410

Страница 1: ...USER MANUAL CRT410 CRT640 CRT900 CRT1250 CONTROLLED ROOM TEMPERATURE CABINETS ...

Страница 2: ...ces or anæsthetic or other flammable gasses Power supply The cabinet is designed for 220 240 V 50 Hz The appliance must be correctly earthed The cabinet is fitted with a B S 1363 UK mains plug A 13 amp fuse must be used The flexible cord fitted has three cores The colours used are Green Yellow Earth Blue Neutral Brown Live Do not use this appliance with multi socket distribution extension leads Ex...

Страница 3: ... that reaches extremes of cold for example in goods bay where the temperature is largely affected by the outside may affect the internal temperature regulation Important Before it can be turned on for the first time the transport safety strap must be removed from the compressor Access the compressor compartment from behind the unit The strap is held on by two cross headed screws at either end see ...

Страница 4: ...ay will flash followed by the highest recorded temperature If is pressed again the display will flash followed by the current chamber temperature Pressing again will return to and the menu will cycle as described above If the buttons are not touched for 10 seconds the screen will return to normal Resetting the Recorded High Low Values Display the value that you would like to reset or as described ...

Страница 5: ... desired from the main screen press P The screen will alternate between and the current set point Use the arrows to adjust it and press P again to confirm Please note that the set point is limited to between 17 and 21 C If the monitored temperature is regularly reaching outside Controlled Room Temperature limits causing high or low alarms please call your supplier for technical assistance Please n...

Страница 6: ... cleaning If it becomes damaged and the cabinet fails to cool please call your supplier for technical assistance Disconnect from the mains power supply before cleaning Monthly use only a light soap solution to clean the internal and external surfaces of the cabinet Do not use any abrasives or cleaning product containing chlorine or other harsh chemicals as they can damage the surfaces and the cool...

Страница 7: ...t changed since before the problem make a note of the details chamber temperature high and low temperatures ambient temperature any alarms etc and call your supplier for technical assistance If the cooling system becomes damaged and the cabinet fails to cool disconnect the appliance from the power supply and call your supplier for technical assistance All work on or repairs to the cooling system m...

Страница 8: ...Unit 21 Botany Business Park Macclesfield Road Whaley Bridge High Peak SK23 7DQ T 01663 735494 www qedscientific com MCRTL21 1 ...

Отзывы: