| +1 (866) 203-8397
Pyxis LT-635 User Manual
6
2
Quick 4-20mA Start
Follow the wiring table below to connect the LT-635 sensor to a controller or PLC.
*NOTE*
All Pyxis sensors
provide a passive 4-20mA output signal, they are NOT LOOP POWERED. 24VDC+ power supply and 4-
20mA+ signal are independent of each other in all Pyxis Lab sensors.
Color
Designation
Brown
Power GND
Green
RS-485 C
Yellow
RS-485 B
Gray*
4-20mA -
(Internally connected to power ground
)
Pink
Not Used
Blue
RS-485 A
White
4-20mA + (TSS)
Red
24V+
Silver
PE
*NOTE*
The 24V- power ground (brown) and the 4-20 mA- (gray) return are internally connected. If the
24V- power ground and the 4-20 mA return in the controller are internally connected (non-isolated 4-20mA
input), it is unnecessary to connect the 4-20 mA- (gray wire) to the 4-20 mA negative terminal in the
controller. If a separate DC power supplier other than that from the controller is used, make sure that the
output from the power supply is rated for 22-26 VDC @ 250mA.
Please refer to the table below for proper LT-635 input 4-20mA scaling in your receiving display, controller,
or PLC.
LT-635 Output Scaling (4-20mA)
mA
Value
Units
4
0
g/L
20
30
g/L