Pyxis LT-635 User Manual
15
4.2
Calibration and Diagnosis by uPyxis Desktop APP
Download and install uPyxis Desktop APP on your PC or Laptop from
https://upyxis.pyxis-lab.com.cn/release/pc/uPyxis.Setup-latest.zip
Connect and power the LT-635 sensor to a display, controller, or PLC. Insert the Pyxis MA-CR Bluetooth
adapter (P/N: MA-CR) as shown in the following connection diagram, using the 8-pin adapters on the
provided sensor cables. (Figure 17). Insert the MA-NEB Bluetooth USB adapter (P/N: MA-NEB) into the
USB port of your PC or Laptop according to connection diagram below (Figure 17).
Figure 17 - Connect the LT-635 to a computer via Pyxis MA-CR and MA-NEB adapter
Establish connection between uPyxis Desktop APP and the LT-635 through the following steps:
1.
Open the desktop
uPyxis Desktop
APP on your PC or Laptop.
2.
Click
Device
to launch the connection option menu.
3.
Select
Connect via USB-Bluetooth
(Figure 18).
4.
Select the Comm Port to make a connection (Figure 19) normally only one Comm port is identified
by uPyxis. If more than one Comm port listed in the selection dropdown, you may try to select
each one to see if a connection can be made. Alternatively, you may use the Windows Device
Manager to identify the Comm Port that the Pyxis USB adapter is used.
After the connection is established, the LT-635 sensor series number and current suspended solids
concentration reading of the water will be displayed on the left of the information page (Figure ). In
this page, you can set the
Device Name, Product Name, Modbus Address
and
Brush Setting
for the
device. Click Calibration to launch the calibration (Figure 21).