Pyxis LT-635 User Manual
17
Figure 21 - Calibration Page
4.2.1
Calibration
Deionized Water Calibration Process - Remove the LT-635 sensor and rinse it thoroughly with deionized
water. Repeat this rinse process three times. Wipe the sensor with a dust-free cloth or paper-towel to
confirm that the sensor is free of obvious impurities. Insert the sensor into a beaker. Inject bubble-free
deionized water into the beaker and wait for the sensor reading to stabilize. Click the “Deionized Water
Calibration” button after the reading stabilizes. If the calibration is successful, the interface will return a
message "Calibration Succeeded". If the calibration has failed, the interface will return a message
"Calibration Failed" and the user must repeat the process until calibration success is achieved.
Figure 22 - Deionized Water Calibration