Please Read Before Installation
The following will damage the Air Check Oxygen monitor.
1.
The Air Check O
2
monitor requires
24 VDC regulated power
.
Please Do Not connect
the monitor to any voltage that exceeds
24 Volts DC, or
ANY AC Voltage.
2. Do not power the Air Check Oxygen monitor with the oxygen sensor
unplugged from the main PC board.
Do Not Connect
the O
2
sensor to
the PC board while the monitor is powered. This Will Damage the O
2
sensor.
3. The oxygen sensor cell is matched to the electronics.
Never exchange
the electronics with an oxygen sensor from a different monitor.
4. When testing Sample Draw and Ex type monitors use an On Demand
regulator to expose the span gas to the monitor. (see section 6.1.2)
5. Only expose the monitor to span gas blends that consist of Oxygen
and Nitrogen only.
Do Not expose the monitor to any combustible
gas,
i.e. Methane, Hydrogen, etc. Exposure to combustible span gases
can damage the oxygen zirconium sensor cell.
6. Do not expose the Oxygen monitor to silicone, Freon or corrosive
compounds. They can cause a loss of sensitivity and damage the
sensor.
7. When using the Air Check O
2
Monitor in wash down areas make sure
you orient the filter with the drain opposite the water flow. (see section
3.2.4)
8.
NOTE:
If using longer lengths of sample tubing, the dust/water filter
must be located
at the end of the sample line to work properly.
9.
The Factory Password for entering the menus is
557