PureAire Monitoring Systems, Inc.
17 |
P a g e
Joystick – You must enter the password to enter the reset function. After the password is entered and
accepted, push the joystick in; (enter) to reset the alarms.
Remote Reset – Refer to section 1.2.4. for location on PC board. The alarm relay board has a two-pin
connector for wiring to a remote switch. When connected to a switch, this remote reset will bypass the
joystick and a password will not be needed to reset the alarms.
NOTE:
The oxygen levels must recover above the alarm thresholds before the horn can be
reset from the remote reset switch or joystick.