PureAire Monitoring Systems, Inc.
41 |
P a g e
Push the joystick right to access the sensor span. The display will indicate a value
between 0 and 255 counts. Pushing the joystick up increases the counts and decreases
the percent oxygen value displayed on the Air Check. Pushing the joystick down
decreases the counts and increases the oxygen value displayed on the Air Check. As the
counts increase and decrease the percent oxygen displayed will also increase and
decrease. Adjust the digital display until 20.9% + / - 0.1% is displayed. The actual
oxygen reading will fluctuate from 20.8% to 21.0%.
Press
ENTER
to accept this value. The digital display will revert to
Set Sensor
Span.
Press the joystick to the left to revert to the
Sensor Adjustment
menu.
NOTE: The “Set Module Zero” menu is not available for the Oxygen monitor
This menu was designed for PureAire toxic and corrosive gas
monitors. When selected, nothing will happen. Push the joystick
Left to leave this menu.
5.5.11 Main Operation Mode
To select the main menu from any sub menu, push the joystick left until the Main Menu
appears. The digital display will indicate the following:
..Set Sensor Span..
20.9%
093
20.9%
..Set Sensor Span..
20.9%
..Sensor Adjustment..
20.9%
Oxygen
20.9%