Page 12
3 Wire
Terminal 1 + ve Current Output (4 – 20mA SINK MODE)
(NOT USED IN SOURCE MODE)
Terminal 2 - ve ground (4-20mA SINK MODE)
Terminal 3 - ve: Direct Current (DC) input (11-30VDC)
Terminal4
+ ve: Direct Current (DC) input (11-30VDC)
Terminal 5 + ve Current Output (4 – 20mA SOURCE MODE)
Terminal
+ ve: 0-10V output (P842 = 1) or 0-5V output (P842 = 0)
Terminal 7 - ve: 0-10V output (P842 = 1) or 0-5V output (P842 = 0)
Terminal 8 Relay 1 Normally Open (N/C) Contact (De-energised)
Terminal 9 Relay 1 Common
Terminal 10 Relay 1 Normally Closed (N/O) Contact (De-energised)
Terminal 11 Relay 2 Normally Open (N/C) Contact (De-energised)
Terminal 12 Relay 2 Common
Terminal 13 Relay 2 Normally Closed (N/O) Contact (De-energised)
Important Information
If the equipment is installed or used in a manner not specified in this
manual, then the protection provided by the equipment may be impaired.
Preparation for Operation
Before switching on, check the following:
The
IMP+
IMP+
IMP+
IMP+
is mounted correctly.
The power supply is correctly installed.