Guidance and manufacture’s declaration – electromagnetic
immunity – for EQUIPMENT and SYSTEMS that are not LIFE-
SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The
CMS60D Pulse Oximeter
is intended for use in the
electromagnetic environment specified below. The customer or the
user of
CMS60D Pulse Oximeter
should assure that it is used in such
an environment.
Immunity
test
IEC
60601
test level
Compliance
level
Electromagnetic
environment -
guidance
Conducted
RF
IEC 61000-
4-6
Radiated
RF
IEC 61000-
4-3
3 V
rms
150 kHz
to 80
MHz
3 V/m
80 MHz
to 2.5
GHz
3 V
rms
3 V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the
CMS60D
Pulse Oximeter,
including cables, than
the recommended
separation distance
calculated from the
equation applicable
to the frequency of
the transmitter.
Recommended
separation distance
P
V
d
1
5
.
3
80 MHz to 800 MHz
IEC 61000-4-
8
characteristic of
a typical
location in a
typical
commercial or
hospital
environment.