Appendix 1
State
Alarm condition
delay
Alarm signal
generation delay
Low voltage alarm
1s
20ms
SpO
2
alarm
330ms
20ms
Pulse rate alarm
330ms
20ms
Probe error alarm
16ms
20ms
Appendix 2
Guidance and manufacture's declaration
Guidance and manufacture’s declaration – electromagnetic
emissions- for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The CMS60D Pulse Oximeter is intended for use in the
electromagnetic environment specified below. The customer of the
user of the CMS60D Pulse Oximeter should assure that it is used in
such and environment.
Emission test
Compliance
Electromagnetic environment
– guidance
RF emissions
CISPR 11
Group 1
The CMS60D Pulse Oximeter
uses RF energy only for its
internal function. Therefore,
its RF emissions are very low
and are not likely to cause any
interference in nearby
electronic equipment.
RF emission
CISPR 11
Class B
The CMS60D Pulse Oximeter is
suitable for use in all
establishments, including
domestic establishments and
those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-
2
Not
applicable
Voltage
fluctuations/
flicker
emissions
Not
applicable