Appendix
Guidance and manufacture’s declaration-electromagnetic emission
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration –electromagnetic emission
The
CMS50D Pulse Oximeter
is tended for use in the electromagnetic environment specified below. The
customer of the user of the
CMS50D Pulse Oximeter
should assure that it is used in such an
environment.
Emission test
compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The
CMS50D Pulse Oximeter
uses RF energy only for
their internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The
CMS50D Pulse Oximeter
is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emission
IEC 61000-3-3
Not applicable
Guidance and manufacture’s declaration-electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration-electromagnetic immunity
The
CMS50D Pulse Oximeter
is intended for use in the electromagnetic environment specified specified
below. The the user of
CMS50D Pulse Oximeter
should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Compliance level
Electromagnetic environment-
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6KV contact
±8KV air
±6KV contact
±8KV air
Floors should be wood, concrete
or ceramic tile. If floor are covered
with synthetic material, the
relative humidity should be at
least 30%.
Power frequency
(50Hz) magnetic
field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment
Guidance and manufacture’s declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration-electromagnetic immunity
The
CMS50D Pulse Oximeter
is intended for use in the electromagnetic environment specified below.
The customer or the user of
CMS50D Pulse Oximeter
should assure that it is used in such an
environment.
Immunity
test
IEC60601 test
level
Compliance
level
Electromagnetic environment -guidance
Radiated
RF
ICE 61000-
4-3
3V/m
80MHz to
2.5GHz
3V/m
Portable and mobile RF communication equipment should
be used no closer to any part of the
CMS50D Pulse
Oximeter
, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
recommended separation distance
P
E
d
=
1
5
.
3
80MHz to 800MHz
P
E
d
=
1
7
800MHz to 2.5GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter