3.3 Clinical Restrictions
1.
As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of subject is
required. For a subject with weak pulse due to shock, low ambient/body temperature, major
bleeding, or use of vascular contracting drug, the SpO
2
waveform (PLETH) will decrease. In this case,
the measurement will be more sensitive to interference.
2.
For those with a substantial amount of staining dilution drug (such as methylene blue, indigo green
and acid indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb) or
thiosalicylic hemoglobin, and some with icterus problem, the SpO
2
determination by this monitor
may be inaccurate.
3.
The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor
blamed for serious error of SpO
2
measure.
4.
As the SpO
2
value serves as a reference value for judgement of anemic anoxia and toxic anoxia,
some patients with serious anemia may also report good SpO
2
measurement.
4. Technical
Specifications
A.
Display Format
: OLED Display
SpO2
Measuring range: 0 ~100 %
Pulse
Measuring range: 30bpm ~250bpm
Perfusion Index Measuring Range:
0%
~
20%
Pulse Wave Display:
columniation display and the waveform display
B.
Power Requirements
: 2 ×1.5V AAA alkaline battery (or using the rechargeable battery instead),
adaptable range: 2.6V~3.6V.
C.
Power Consumption
: Smaller than 30mA
D.
Resolution
: SpO
2
: 1%, Pulse: 1bpm
,
Perfusion Index: 0.1%.
E.
Measurement Accuracy
: ±2% in stage of 70%
-
100% SpO2, and meaningless when stage
being
smaller than 70%. ± 2 bpm during the pulse rate range of 30
-
99 bpm and ± 2% during the pulse rate
range of 100~250 bpm
F.
Measurement Performance in Weak Filling Condition
: SpO2 and pulse rate can be shown correctly
when pulse-
filling ratio is 0.4%. SpO2 error is ±4%, pulse rate error is ±2 bpm or ±2% (select larger).
G.
Resistance to surrounding light:
The deviation between the value measured in the condition of
man-
made light or indoor natural light and that of darkroom is less than ±1%.
H.
It is equipped with a function switch
. The Oximeter can be powered off in case no finger is the
Oximeter within 5 seconds.
I.
Optical Sensor
Red light (wavelength is 660nm, 6.65mW)
Infrared (wavelength is 880nm,6.75mW)
5. Accessories
Hanging rope, pouch, silicone protector, user manual
6. Installation
6.1 View of the front panel
Fig. 2 Front view
Fig. 3 Batteries installation
6.2 Battery
Step 1. Refer to Figure 3 and insert the two AAA size batteries properly in the
right direction.
Step 2. Replace the cover.
Please take care when you insert the batteries for the improper insertion
may damage the device.