Please check the packing before use to make sure the device and accessories are totally in
accordance with the packing list, or else the device may have the possibility of working abnormally.
!
Please don't measure this device with functional tester for the device's related information.
1.3 Attention
•
Keep the oximeter away from dust, vibration, corrosive substances, explosive materials, high
temperature and moisture.
•
If the oximeter gets wet, please stop operating it.
•
When it is carried from cold environment to warm or humid environment, please do not use it
immediately.
•
DO NOT operate keys on front panel with sharp materials.
•
High temperature or high pressure steam disinfection of the oximeter is not permitted. Refer to
User Manual in the relative chapter for instructions of cleaning and disinfection.
•
Do not immerse the oximeter in liquid. When it needs cleaning, please wipe its surface with
disinfectant solution using a soft cloth. Do not spray any liquid directly onto the device.
•
When cleaning the device with water, the temperature should be lower than 60°C.
•
As to the fingers which are too thin or too cold, it would probably affect the normal measure of the
patients' SpO2 and pulse rate, please clip the thick finger such as thumb and middle finger deeply
enough into the probe.
•
Do not use the device on infant or neonatal patients.
•
The product is suitable for children above four years old and adults (Weight should be between 15
kg to 110 kg).
•
The device may not work for all patients. If you are unable to achieve stable readings, discontinue
use.
•
The update period of data is less than 5 seconds, which is changeable according to different
individual pulse rate.
•
Please read the measured value when the waveform on screen is equably and steady-going. This
measured value is optimal value. And the waveform at the moment is the standard one.
•
If some abnormal conditions appear on the screen during test process, pull out the finger and
reinsert to restore normal use.
•
The device has normal useful life for three years since the first electrified use.
•
The hanging rope attached to the device is made from Non- allergy material, if particular group are
sensitive to the hanging rope, stop using it. In addition, pay attention to the use of the hanging rope,
do not wear it around the neck on the purpose of avoiding harm to the patient.
•
The instrument does not have low-voltage alarm function, it only shows the low-voltage. Please
change the battery when the battery energy is used out.
•
The instrument dose not have alarm function.Do not use the device in situations where alarms are
required.
•
Batteries must be removed if the device is going to be stored for more than one month, or else
batteries may leak.
•
A flexible circuit connects the two parts of the device. Do not twist or pull on the connection.
1.4 Indication for Use
The Fingertip Pulse Oximeter is a non-invasive device intended for the spot-check of oxygen saturation
of arterial hemoglobin (SpO
2
) and the pulse rate of adult and pediatric patients in home and hospital
environments (including clinical use in internist/surgery, anesthesia, intensive care ect.). This device is
not intended for continuous monitoring.
2. Overview
The pulse oxygen saturation is the percentage of HbO
2
in the total Hb in the blood, so-called the O
2
concentration in the blood. It is an important bio-parameter for the respiration. For the purpose of
measuring the SpO
2
more easily and accurately, our company developed the Pulse Oximeter. At the
same time, the device can measure the pulse rate simultaneously.
The Pulse Oximeter features in small volume, low power consumption, convenient operation and being
portable. It is only necessary for patient to put one of his fingers into a fingertip photoelectric sensor for
diagnosis, and a display screen will directly show measured value of Hemoglobin Saturation.
2.1 Classification
Class II b, (MDD93/42/EEC IX Rule 10)