HL Series – Instructions manual
34
TROUBLE SHOOTING
ERROR
DESCRIPTION
ACTION TO CARRY OUT
EFFECT
SOLUTION
Not stable weight
: The
visualized weight value
keeps going up and
down.
Weight drift:
The
visualized weight keeps
increasing and
decreasing,
POINT 1:
press T/CLEAR and then
ENTER.
The message ERROR
disappears and the
weight becomes stable
The scale was out of range; by pressing the key
T/CLEAR the problem was solved through the
scale software.
Either the message
ERROR is still shown
on the display, or the
weight is visualized in
an unstable way.
Check the integrity of the weighing system and
of the load cells. To better understand what is
broken, carry out the action described in the
POINT 2.
POINT 2:
Disconnect the 5-pin
connector (metal circular connector)
of the load cell cable from the scale.
Weight still fluctuates
/drifts / shows ERROR.
Check if the 5-pin female connector of the scale
has humid contacts. If so, use warm air to dry
off the connector (use for instance a hairdryer).
If after this operation the weight still fluctuates,
contact service.
Weight gets stable,
even though a wrong
value is shown.
Carry out the action described in the
POINT 3.
POINT 3:
Disconnect the junction
cable from all the load cells and
reconnect the 5-pin connector to the
scale, avoiding the contact between
the different connection wires.
Weight still fluctuates
/drifts / shows ERROR.
Check if the cable is damaged; if so, replace it.
Check if the junction box contains humidity; if
so, dry the junction box.
Weight gets stable.
The reason of the problem is that one or more
load cells do not work properly. To better
understand the cause of the failure, carry out
the action described in the
POINT 4
POINT 4:
press T/CLEAR and after
ENTER
Wait for some seconds. Connect one
single load cell to the junction cable,
by connecting each single wire colour
by colour, and pay attention that the
wires do not touch each other.
Weight still fluctuates
/drifts / shows ERROR.
The connected load cell doesn't work properly;
contact service.
Weight gets stable.
The load cell doesn't work properly; disconnect
the load cell and connect another one. This
operation must be repeated for all load cells.
Weight fluctuates, or it is
not stable while working
only.
POINT 5:
Make sure that the cables
of the load cells are not torn or
crushed.
The cables are
seriously damaged or
broken.
Replace the damaged cables.
The cables are intact
Carry out the action described in
POINT 6
POINT 6
: Check that the tank, or the
device receiving the weight, does not
touch the frame in any point.
The tank lays on the
frame.
Arrange for the tank not to touch the frame in
any point.
The tank does not lay
on the frame.
Contact service.