background image

3

Thread hole for accelerometer VIB 6.195

 

Dimensions in mm

17

12

Ø 27

M8

90°

0,5±0,2

• Clean the contact surfaces of the accelerometer 

/ adapter and of the machine with solvent.

• Allow the contact surfaces to dry.
• Apply a thin film of threadlocker (LOCTITE 243) 

to improve signal transmission.

• Screw in the accelerometer / adapter and tight-

en with a wrench.

Maximum torque:

 

– 11 Nm, accelerometer VIB 6.195 w/ M8 thread

 

– 11 Nm, adapter VIB 3.437

 

– 3 Nm, adapter VIB 3.439

• Screw in the accelerometer by hand (3 to 7 Nm).

Notes

Excessive torque can damage the threads on the 
accelerometer or machine part. Too little torque 
can allow the accelerometer to work loose and 
cause measurement errors!

If the accelerometer is affixed to non-grounded 
machine components (e.g. belt-driven fan), the 
accelerometer must be grounded to prevent static 
charges.

Adhesive mounting

Applies to bonded adapter VIB 3.433.

Required tools and resources

 

– Hand-held drill and drill bit (3.5 mm)

 

– Angle grinder and rasp

 

– Compressed air for cleaning out the hole.

 

– Torque wrench with 22 mm / 7/8“ hex socket

 

– Solvent for degreasing

 

– Threadlocker (LOCTITE 243)

 

– 2-Component adhesive (WEICON HB 300,..)

Note

Switch off the machine and secure it against a 
restart. Do not switch on the machine until the 
adhesive has hardened (approx. 24 hours).

Select the point of installation

• Allow sufficient space for applying the adhesive 

with a wooden spatula.

Prepare bonding location

• Grind down existing coats to the bare metal (Ø 

> 30 mm, 

A

). 

• If necessary, grind down the location (

A

).

• Sand down the point of installation with a rasp 

and file several grooves in a diamond pattern 
for greater adhesive strength (

A

).

Optionally, and only if drilling is possible: 

• Drill hole for centering pin: 3.6 mm / 5 mm deep 

(

B

). Remove the centering pin from the mount-

ing adapter if drilling is not possible.

• Blow out the hole.
• Clean the contact surfaces of the mounting 

adapter and the machine with solvent.

• Allow contact surfaces to dry.

30

3,6

5

Dimensions in mm

A

B

C

D

Содержание VIB 6.195

Страница 1: ...Accelerometer Wind VIB 6 195 Installation and Operation ...

Страница 2: ...and Maintenance The operation of the accelerometer is mainte nance free A repair is not possible Disposal Dispose of the accelerometer after use in accor dance with applicable national regulations Mounting The frequency response and dynamic range of an accelerometer can be greatly influenced by the installation Poor coupling to the measurement location can adversely affect the signal and restrict ...

Страница 3: ...equired tools and resources Hand held drill and drill bit 3 5 mm Angle grinder and rasp Compressed air for cleaning out the hole Torque wrench with 22 mm 7 8 hex socket Solvent for degreasing Threadlocker LOCTITE 243 2 Component adhesive WEICON HB 300 Note Switch off the machine and secure it against a restart Do not switch on the machine until the adhesive has hardened approx 24 hours Select the ...

Страница 4: ...o cation and turn it to evenly distribute the adhe sive If applicable screw in the centering pin D Do not remove excess adhesive If necessary apply additional adhesive around the bonding location for greater stability Note An adhesive tape may be used to hold the adapter in place during the first hour of hardening Mount sensor Clean the contact surfaces of the sensor and the adapter with solvent A...

Страница 5: ...7 18 VDC Noise 0 1 Hz 20 kHz 0 002 ms 2 as of 2 Hz Output impedance 10 Ohm Environment Degree of protection IP 67 w cable connected Shock resistance 50 kms 2 5000 g Temperature range 30 C 80 C 22 F 176 F Temperature sensitivity 0 01ms 2 K Transverse sensitivity 5 Magnetic sensitivity 1ms 2 T at 50 Hz Base strain sensitivity 0 1ms 2 µm m Mechanical Case material Stainless steel VA 1 4305 Connector ...

Страница 6: ... cable Lay the cable Lay the cable in a cable conduit or a protective tube Use Velcro strips or cable ties for mounting Do not lay the connection cable parallel to pow er lines Keep a minimum distance 1m Lay a cable loop with sufficient excess cord at each connection or before each gland Label the cable ends to avoid confusion Observe the terminal assignments on the CMS see CMS installation manual...

Страница 7: ...5 449 CLD VIB 5 436 VIB 311221 10 TNC VIB 5 449 CLD VIBSCANNER VIBXPERT II VIBGUARD portable VIB 5 444 5 Mini Snap Extension cable optional Cable adapter Connection options for accelerometer VIB 6 195 to portable measurement devices ...

Страница 8: ...PRÜFTECHNIK 85737 Ismaning Germany www pruftechnik com LIT 61 201 EN 08 2018 Productive maintenance technology ...