Page 61 of 69
Proton Products InteliSENS DG2030 Diameter Gauge Instruction Manual (Issue 1a)
DW
Bit
Comments
Unit
Range/Remark
Default value
DW25
DW26
Preset Z lower limit
1=0.01um/0.01mil
0~20000000
50000
DW27
DW28
Preset Ovality upper limit
1=0.01um/0.01mil
0~20000000
50000
DW29
DW30
Preset Ovality lower limit
1=0.01um/0.01mil
0~20000000
50000
DW31
DW32
Preset SMFD upper limit
1=0.01um/0.01mil/
1=0.1%
0~20000000
50000
DW33
DW34
Preset SMFD lower limit
1=0.01um/0.01mil/
1=0.1%
0~20000000
50000
DW35
DW36
Preset core diameter
1=0.01um/0.01mil
0~20000000
800000
DW37
DW38
Diameter Average time
1=1ms
0~5000
1000
DW39
Shrinkage
1=1um{0.1mil}/
1=0.1%
0~10000
0
DW40
Pitch for helix
1=10um{1=1mil)
1~65535
1000
DW41
Core number for helix
1=1
1~100
2
DW42
Flaw interval time
1=0.001s/1=0.001m{1
=0.001ft}
1~65535
100
DW43
SMFD reference diameter
average time
1=1ms
1~1000
100
DW44
SMFD measurement
average time
1=0.1ms
1~100
10
DW45
Start speed for SMFD
detection
1=1m/min{1=1ft/min}
1~65535
0
DW46
Start Length for SMFD
1=1m{ft}
0
DW47
Start Length for Diameter
over tolerance
1=1m{ft}
0
DW48
Reset length / running
maximum and minimum
/flaw counting
1=Reset;other=Invalid
0
DW49
Logic input
0
Logic input 1 function
0=Reset;1=External
Alarm 1;2=External
Alarm 2;3=Line rev
direction
0
1
2
Logic input 2 function
0=Reset;1=External
Alarm 1;2=External
Alarm 2;3=Line rev
direction
1
3
4
The polarity of logic input 1
0=Active low; 1=Active
high
0
5
The polarity of logic input 2
0=Active low; 1=Active
high
0
DW50
Relays
0
Relay 1 function
0=Gauge OK;1=Over
all limit; 2=Under all
limit;3=Flaw;4=Lump;5
=Neck, 6=Over
Average Upper
0
1
2