E
THER
N
ET
/
IP
COMMUNICATIONS
The EtherNet / IP interface is an optional extra that must be ordered for installation during
manufacture; it cannot be retrofitted to the gauge.
The Ethernet/IP interface is compatible with PLCs fitted with an Ethernet/IP interface that supports the
CIP (Common Industrial Protocol).
This communications interface operates independently of the other communications interfaces and
may be used at the same time as them.
EtherNet / IP interface
The EtherNet / IP interface may be accessed through the “PROFI/EIP/DEV” connector.
Connector type: DB9 female (socket)
Pin EtherNet/IP Comments
1
LAN TX-
2
LAN TX+
7
LAN RX-
9
LAN RX+
S
Shield
Ensure that the cable shield is connected to this via the plug shield connection.
EtherNet / IP LED indicator
i-BUS
LED status
Indication
Continuous green
Online
EtherNet / IP communications
Continuous red
Communication error
Extinguished
No communication