Page 16 of 68
Proton Products InteliSENS DG1130 Diameter Gauge Instruction Manual (Issue 1a)
O
PTICAL WINDOW CLEANING PROCEDURE
•
The optical windows are manufactured from anti-reflection coated optical glass; they must be
treated with the same level of care as a high-performance camera lens.
•
Before inspecting or cleaning the optical windows, ensure that the gauge is powered
off and no light is emitted.
Required items
Notes
Small blower brush
Such as the type used to remove dust from camera lenses.
Lens cleaning tissues or
micro-fibre lens cleaning cloth
Do NOT use facial tissues as these can scratch delicate optics.
Lens cleaning solution
Such as the type specified for cleaning camera lenses.
1. Use the small blower brush to remove any visible dust on the optical window.
2. Apply a few drops of lens cleaning solution to a fresh lens cleaning tissue or a clean micro-
fibre lens cleaning cloth.
3. Gently wipe the optical window from the centre outwards; apply only light pressure to the
tissue or cloth when wiping the optical window.
4. Repeat as necessary with fresh tissues or a clean section of cloth until the optical window is
clean and free of all smears and smudges.
I
NSTALLATION SEQUENCE
Unpack the gauge and check for missing accessories and shipping damage.
Mechanical installation:
1. Mount the gauge securely on a user supplied platform using the four mounting brackets.
Electrical installation:
1. Install earth connections.
2. Install optional CDI4 interface display unit.
3. Depending on the model, install either an AC power cable or DC power supply (such as an
optional Proton Products PSU-UNI, PSU-BOB or PSU-CAN power supply).
4. Install interface connections (RS-232, RS-422/485, Ethernet or optional PROFIBUS,
EtherNet/IP or WiFi antenna).
5. Install speed input (analogue or pulse).
6. Install optional electrical interfaces (Logic inputs, Relay outputs, Analogue outputs, PI
feedback controller connections) using the optional Proton Products PSU-BOB breakout box
or terminal strip.
7. Configure the gauge either via an optional CDI4 interface display unit or any of the above
interfaces.