Propaq CS Directions for Use
101
Ref
erence
Real-Time ECG Analog/Defib Sync Specifications
Special cables are required to interface the defib sync connector to a Physio-Control LIFEPAK 5 or
LIFEPAK 6s defibrillator. The sync and real-time ECG outputs do not operate during in-service mode.
Signal
Specification
Sync Output
0 to 5 V pulse, 100 ±5 ms wide, starts within 35 ms after peak of
R-wave. 15 mA short circuit current.
Real-time ECG Output
Range = ±6 V minimum, centered about 0 V, Gain = 1000X,
noninverting for lead II, inverting for all other leads, delay <3 ms,
0.05-100 Hz, going to -5.9 V ±5% during ECG lead fail. V lead has
no Real-Time analog output.
Marker Input (Defib Sync only)
Normally 0 V in, a pulse either ±3 to ±15 V for 10-70 ms puts a
marker in ECG trace. ~ 5 k
Ω
input resistance.
Shield
Common terminal for other signals
N/C
N/C
GROUND
MARKER IN
SYNC
OUT
N/C
Real-time ECG Output connector side panel view
Defib Sync connector side panel view
ECG
N/C
GROUND
SHIELD
N/C
SHIELD
Содержание Propaq 242
Страница 6: ...6 Protocol Systems Inc...
Страница 18: ...18 Protocol Systems Inc...
Страница 26: ...26 Protocol Systems Inc...
Страница 50: ...50 Protocol Systems Inc...
Страница 52: ...52 Protocol Systems Inc...
Страница 64: ...64 Protocol Systems Inc...
Страница 66: ...66 Protocol Systems Inc...
Страница 72: ...72 Protocol Systems Inc...
Страница 74: ...74 Protocol Systems Inc...
Страница 78: ...78 Protocol Systems Inc...
Страница 84: ...84 Protocol Systems Inc...
Страница 86: ...86 Protocol Systems Inc...
Страница 94: ...94 Protocol Systems Inc...
Страница 120: ...120 Protocol Systems Inc...