Propaq CS Directions for Use
89
Maintenance
Replace Monitor Input Power Fuse
If the green battery charging light is off and the ac power adapter does not provide power to the
monitor even when all connections are intact, the monitor’s input power fuse may need to be
replaced. This procedure must be performed by a qualified service person. To change fuses:
1.
Disconnect the monitor from the patient and turn
off the monitor.
2.
Disconnect the ac power adapter from the
monitor.
3.
Using a small, flat-blade screwdriver, turn the
fuse carrier counterclockwise to release it.
4.
Remove the fuse carrier and replace the fuse with
the type 3A/250V, 2AG.
DEFIB SYNCHRO
MONITOR
EKG x 1000
!
3A
2AG
!
!
!
12-28V, 3A
Power
input fuse
holder
Green
battery
charging
light
Right Side Panel