The film developer OPTIMAX
®
2010 has been extended by a further unit version.
Introduction
Processing time and developer temperature of the OPTIMAX
®
2010 NDT processor
have been adjusted to the requirements of NDT-films. This results in the following
modifications compared with the OPTIMAX
®
2010 standard version.
General
To maintain constantly good film quality the NDT-roller (upper roller of last roller pair
of developer rack) must be replaced every 3 - 6 months, depending on the quantity
of films processed, the chemicals and films used. 1 spare roller is supplied with the
unit. This roller must be considered a wear part and is excluded from the guarantee.
Technical Specifications (compare page 8)
Processing time (compare page 17)
Processing time from 1.5 - 10.0 min
Developer temperature (compare page 18)
Adjustable 26 - 37 °C / 78 - 99 °F)
X:
1
50 Hz
Y:
0
Version with 1 pump
2
60 Hz
2
Version with 2 pumps
Important!
Please take care that the film does not carry any dirt into the
processor.
Processing capacity:
129 films 24x30 cm (crosswise), at lowest
processing time
Process time:
1.5 - 10 min
Linear speed:
8.4 - 56.0 cm/min
Developer time:
25-164 s
Developer temperature
26 - 37 °C / 78 - 99 °F
Processing and developer time relation
Processing time (min)
Developer time (s)
Infeed speed (cm/min)
1,5
25
56
1,7
28
49
2,0
33
42
2,3
38
37
2,5
41
34
2,7
44
31
3,0
49
28
5,0
82
17
8,0
131
11
10,0
164
9
Содержание optimax 2010
Страница 28: ...32 OPTIMAX 2010 Accessories...
Страница 45: ...OPTIMAX 2010 49 Electric Diagrams English...
Страница 46: ...50 OPTIMAX 2010 Electric Diagrams...
Страница 47: ...OPTIMAX 2010 51 Electric Diagrams English...
Страница 48: ...52 OPTIMAX 2010 Electric Diagrams...
Страница 52: ...56 OPTIMAX 2010...