34
OPTIMAX
®
2010
Installation Data
Installation Data
Optional worktable or
closed cabinet (accesso-
ries)
Measures and positions are recommendations
1.
Wall socket
220-240 V, 16 A resp. 110 V,
15 A (depending on machine
model). Power lead should be
equipped with Earth-Leakage
Switch, 25 A / 30 mA nominal
error-current. In addition, a
power control switch can be
installed.
2.
Fresh water connection 3/4”
with stop cock, permissible
pressure 2-10 bar, water tem-
perature 5-25 °C.
3.
Drainage plastic pipe PVC
∅
50 mm (2“) incl. syphon.
4.
Drainage resp. collecting con-
tainers for used developer.
5.
Drainage resp. collecting con-
tainers for used fixer.
6.
Storing space for replenish-
ment tanks: Below machine or
externally.
7.
Ventilation of darkroom is nec-
essary.
8.
Sink with freshwater and flexi-
ble hose. Interior dimensions
minimum (LxWxH)
60x40x30 cm (24x16x12“).
Darkroom
Sink
60 x 40 cm (24 x 16“)
1
2
3
4 5
6
7
8