Trouble Shooting
6
Replenishment
6.1 Replenishment pump does not work or works insufficiently
7
Dryer
7.1 Dryer fan does not run or runs with too low speed
•
Check the correct connecting of the fan cables:
bl = blue; bk = black; br =
brown
.
•
If the fan supply lines have been transposed, the dryer fan will run at half the
nominal output only.
8
Water
8.1 Rinsing water does not flow
•
The water pressure in the water supply system is too low: Minimum pressure
2 bar (29 psi).
•
The valve opens, but no water flows through: filter at inflow is blocked.
•
Check the green water inlet hose inside the machine.
•
The water level switch is not positioned correctly: switching point is too low
(below the overflow level), therefore no fresh water is supplied.
•
The level switch is defective (e. g. permanently open).
•
Start the unit in manual mode and measure the voltage at the connection X4.
If no voltage can be registered, replace the power output section PCB.
8.2 Water tank overflows
•
Water drainage hose (overflow) should have a constant fall. The hose end
should be positioned above the drainage level in the syphon.›‹›‹
•
Check water drain opening in the tank and the hose for blockage and resi-
dues.
•
When extreme algae growth is noticed, the overflow can be connected
directly at the back of the water tank.
8.3 Rinsing water constantly switches on and off
•
At regular intervals (approx. 1 minute on and 1 minute off): The water saving
mode has been activated.
•
At short intervals: th level switch is not positioned correctly (exactly on the
overflow level). Insert the rack. If the error continues to occur, readjust the
height of the level switch.
•
Check filter in the suction pipe (repl. container) and clean it if necessary.
•
Replenishment hose aspirates air. Check connections.
•
Check eccentric position. Flow rate is 240 ml/min at maximum eccenter posi-
tion 100%.
•
(60 Hz: 240 ml/min at 85 %!)
•
Measure the voltage at the connection X6 (replenishment pump) on the power
output section while the replenishment is switched on (in manual mode). If no
voltage can be registered - replace the power section PCB. Remark: 2 pump
version: developer replenishment pump at X6 and fixer replenishment pump
at X5.
•
Clean valves inside connection tube of pump.
Install valve-insert correctly: Pay attention to flow-through di-
rection!
Содержание optimax 2010
Страница 28: ...32 OPTIMAX 2010 Accessories...
Страница 45: ...OPTIMAX 2010 49 Electric Diagrams English...
Страница 46: ...50 OPTIMAX 2010 Electric Diagrams...
Страница 47: ...OPTIMAX 2010 51 Electric Diagrams English...
Страница 48: ...52 OPTIMAX 2010 Electric Diagrams...
Страница 52: ...56 OPTIMAX 2010...