Spare Parts
E
n
g
li
s
h
Tips and Tricks
Removal of control PCB
To reach the screws of the control PCB, remove the film covering the buttons by
approx 20 mm on the top and bottom edges (lift it, see illustration top left).
Stop start-cycle
The start-cycle (after switching the machine on) can be manually interrupted. To
stop start-cycle, press both arrow-buttons (2+3) simultaneously. The start-cycle
may only be interrupted for service purposes.
Display of machine information
When during the start-up cycle one of the arrow-buttons is pressed then various
machine information will be displayed.
Arrow button “Up”:
The software version is displayed as long as you press the button.
Arrow button “Down”:
The number of film cycles is displayed as long as you press the button. Not the dec-
imal point. Since the display has two digits only, the value needs to be converted:
XX : value x 10 = number of films
XX. : value x 100 = number of films
X.X : value x 1,000 = number of films
X.X. :value x 10,000= number of films
9.9. : more than 990,000 films
Pos.
Order No.
Description:
130
1160-0-3103
Indexing bolt
131
1160-0-3106
Blind for light protection cover
132
2011-0-0137
Control panel foil
-
2011-0-0139
Operating membrane, 2 pump version
133
0160-9-1200
Control panel
132
133
Once the operating panel PCB has been
removed, the control unit can be
removed. Please do not lift the control unit
out further than shown in the illustration. A
guide on the top edge allows to position
the control unit for service purposes.
Attention!
The heat sink is not
earthed!
Guide slot
Содержание optimax 2010
Страница 28: ...32 OPTIMAX 2010 Accessories...
Страница 45: ...OPTIMAX 2010 49 Electric Diagrams English...
Страница 46: ...50 OPTIMAX 2010 Electric Diagrams...
Страница 47: ...OPTIMAX 2010 51 Electric Diagrams English...
Страница 48: ...52 OPTIMAX 2010 Electric Diagrams...
Страница 52: ...56 OPTIMAX 2010...