(ii) Changing from LEV to Standard Configuration
Promega Corporation ·
2800 Woods Hollow Road
·
Madison, WI 53711-5399 USA
·
Toll Free in USA 800-356-9526
·
Phone 608-274-4330
·
Fax 608-277-2516
· www.promega.com
Part# TM300
Printed in USA.
Page 32
2/08
1.
Ensure that the power is turned off and the instrument is unplugged. This will release the motors so that
the heads can be moved gently to allow easier access to the hardware assemblies.
2.
Gently and slowly, with constant pressure on both the right and left sides, push the black LEV Magnetic
Rod Assembly down to the lowest position.
Do not push too fast
, as this could result in damage to the
instrument’s electronics.
3.
Unscrew the three thumbscrews on top of the magnetic
rod assembly.
4.
Gently lift up the LEV Magnetic Rod
Assembly to remove it.
5.
Pull down on the LEV Plunger Bar Adaptor to
release the magnets that hold it in place.
Hint:
Pull down at an angle so that the
back of the magnet is released first.
Remove the black LEV Plunger Bar Adaptor.
6.
Insert the blue SEV Magnetic Rod Assembly.
7.
Secure the Magnetic Rod Assembly with the three thumbscrews, and hand-tighten.
8.
Complete the configuration by updating the Firmware to the SEV setting (Section VIII.A).
Failure to change the instrument firmware operational mode from LEV to SEV will cause damage to
the instrument.
LEV Magnetic Rod Assembly
LEV Plunger Bar Adaptor
SEV Magnetic Rod Assembly
!
Thumbscrew
tm300.0208_Corrected.qxp 3/11/2008 10:43 AM Page 32