5. If the hardware mode in the firmware has changed, the following LCD screens will appear,
prompting the user to change the hardware. Failure to change the hardware will result in
damage to the instrument. Directions for changing the hardware are provided in Section
VIII.B.
If changing from LEV to SEV, the following screens will appear:
6214MB
==Hardware Setup==
Remove LEV hardware:
Magnet & Plunger Bar
Press Run/Stop Key
==Hardware Setup==
Add SEV hardware:
Magnet Assembly
Press Run/Stop Key
If changing from SEV to LEV, the following screens will appear:
6220MB
==Hardware Setup==
Remove SEV hardware:
Magnet Assembly
Press Run/Stop Key
==Hardware Setup==
Add LEV hardware:
Magnet & Plunger Bar
Press Run/Stop Key
6. The display will show the operational mode selected.
Turn the instrument off and then on to cycle the
instrument power.
Note:
The selected operational mode or hardware
setting will not be applied until you cycle the power.
60
3
4MB
Forensic Mode Set
LEV Hardware
Please Cycle Power
==Setup Verification==
7. When you turn the instrument back on, verify the
instrument’s operational mode setting on the display
screen.
Fnsc
Promega MAXWELL 16
Purification System
Version 4.0
LEV
Operational Mode Setting
Rsch = Research
Fnsc = Forensic
IVD = Clinical
SEV = Standard Elution Volume
LEV = Low Elution Volume
tm300.0208_Corrected.qxp 3/11/2008 10:41 AM Page 29