3.

Operating the Quantus™ Fluorometer 

The Quantus™ Fluorometer uses five buttons for simple navigation [one center 
button to confirm your choice (“Go”), and four surrounding buttons to navigate 
up, down, left or right through the menus]. 

The instrument is designed for use with Promega QuantiFluor

®

 Dye Systems 

for fluorescence-based quantitation of dsDNA, ssDNA and RNA. A protocol for 
each QuantiFluor

®

 Dye is listed on the Protocol screen. Each of these protocols 

contains optimized calibration curve settings based on the blank, standard, 
sample and dye preparation described in Section 3.A. The instrument also may 
be used with other fluorescent dyes capable of excitation and emission in the 
proper wavelength range. If you are using a dye other than one of the 
QuantiFluor

®

 Dyes, a different dye dilution or a different standard or unknown 

sample preparation method, follow the instructions in Section 3.B to create a 
user-defined protocol. 

The Quantus™ Fluorometer uses a single-point calibration process. The 
fluorometer must be calibrated prior to the first use of a protocol. If a protocol 
that has never been calibrated is selected, the Calibration screen will appear. If 
a protocol has been calibrated, the Home screen will appear, and you can 
proceed to Section 3.D. Each Quantus™ Fluorometer protocol has calibration 
values associated with the specific assay reagent. Once the protocol of interest 
is calibrated, the Quantus™ Fluorometer will save the calibration values for 
future use and will not require recalibration. However, we recommend 
performing a calibration with each new lot of dye reagent to ensure the most 
accurate measurement. 

Materials to Be Supplied By User 

thin-walled PCR tubes (Cat.# E4942 or Axygen Cat.# PCR-05-C, available from 
Fisher or VWR) 

3.A. Calibrating the Quantus™ Fluorometer for Use With the QuantiFluor

®

 Dye 

Systems 

Follow this protocol to calibrate the Quantus™ Fluorometer prior to 
quantitation using the QuantiFluor

®

 Dye Systems. The table below lists the 

dilution factors used to prepare the standard samples for the various 
QuantiFluor

®

 Dye Systems when using the standard provided with the 

system. If using a different DNA or RNA sample as a standard, prepare the 
standard sample at the final standard calibration (ng/tube) indicated below. 

Blank and Standard Sample Preparation 

1. Prepare 1X TE buffer by diluting 20X TE Buffer (pH 7.5) to 1X with 

nuclease-free water. For example, combine 200µl of 20X TE Buffer and 
3,800µl of nuclease-free water to prepare 4ml of 1X TE buffer. 

Promega Corporation

 

·

 2800 Woods Hollow Road 

·

 Madison, WI 53711-5399 USA  

Toll Free in USA 800-356-9526 

·

 Phone 608-274-4330 

·

 Fax 608-277-2516 

·

 

www.promega.com

 

Part# TM396

Printed in USA. 

Page 8

Revised 1/20

TM396.0120_EIVD_TM.qxd  1/27/2020  2:50 PM  Page 8

Содержание E6150

Страница 1: ...T e c h n i c a l M a n u a l Quantus Fluorometer Operating Manual INSTRUCTIONS FOR USE OF PRODUCT E6150 PRINTED IN USA Revised 1 20 Part TM396 TM396 0120_EIVD_TM qxd 1 27 2020 2 50 PM Page a...

Страница 2: ...shooting 15 A Common Problems 15 B Error Messages 16 6 Appendix 17 A Updating Firmware 17 B Warranty and Service 17 C Certificate of Decontamination 18 D Related Products 19 7 Summary of Change 19 1 I...

Страница 3: ...y Includes Quantus Fluorometer USB Power Supply USB Cord Outlet Adapter Kit 4 adapters supplied for international compatibility Quick Reference Guide Operating Manual Product Registration Card Shippin...

Страница 4: ...on sensor Excitation Peak Red 625nm Blue 470nm Excitation Filters Red 640nm shortpass wavelengths up to 640nm Blue 495nm shortpass wavelengths up to 495nm Emission Filters Red 660 720nm Blue 510 580nm...

Страница 5: ...ques Veuillez contacter votre repr sentant Promega local concernant la destruction de l appareil Veuillez respecter les exigences de votre tablissement concernant la destruction des accessoires 9630TA...

Страница 6: ...nt may be impaired Do not disassemble the unit The Quantus Fluorometer accepts 0 5ml PCR tubes Do not force larger tubes into the tube holder Use caution around solvents because they may corrode the p...

Страница 7: ...etting Up the Quantus Fluorometer 1 Position the Quantus Fluorometer on a flat surface 2 Select the appropriate outlet adapter for your wall outlet from the Outlet Adapter Kit provided Promega Corpora...

Страница 8: ...un setup exe Follow prompts to install the software If NET 4 0 framework is not installed on the computer the software installer will prompt the user to download and install the framework software ava...

Страница 9: ...s Fluorometer protocol has calibration values associated with the specific assay reagent Once the protocol of interest is calibrated the Quantus Fluorometer will save the calibration values for future...

Страница 10: ...0 single concentration standard curve 1 400 RNA Cat E3310 high concentration standard curve 1 400 RNA Cat E3310 low concentration standard curve 1 2 000 ssDNA Cat E3190 high concentration standard cur...

Страница 11: ...electing the desired protocol navigate to the Calibration screen 2 Place the blank sample into the tube holder and close the lid Select Read Blank and the fluorescence in relative fluorescence units R...

Страница 12: ...tered is 100 200 0 5 4 Choose the appropriate fluorescence channel Red Excitation 640nm shortpass emission 660 720nm Blue Excitation 495nm shortpass emission 510 580nm 5 Select Save Note To edit or de...

Страница 13: ...it is within the linear range of the standard 1 Press any button to wake up the Quantus Fluorometer The Home screen will appear displaying the last protocol used the last concentration measurement an...

Страница 14: ...esult history select Clear from the menu list Results can be transferred to a connected computer as the measurements are read using the Quantus Software see Section 2 B The default fields transferred...

Страница 15: ...f additional cleaning is needed wipe with a cloth dampened with a mild detergent remove detergent with a cloth dampened with water and wipe dry 4 Do not use organic solvents they may damage the plasti...

Страница 16: ...ow concentration then use the low concentration standard and that the standard used is similar to the sample i e if the unknown sample is plasmid DNA then use a plasmid DNA standard Sample concentrati...

Страница 17: ...o retain the calibration values Valid on Calibration screen The standard to blank ratio was within the parameters expected by the instrument Press Save to retain the calibration values and proceed to...

Страница 18: ...e update only when instructed by Promega Technical Services Follow the instructions located at www promega com quantusfluorometerfirmware 6 B Warranty and Service The Quantus Fluorometer comes with a...

Страница 19: ...aminated and can be handled without exposing personnel to health hazards Circle the type of material used in the instrument Chemical Biological Radioactive Briefly describe the decontamination procedu...

Страница 20: ...00 356 9526 Phone 608 274 4330 Fax 608 277 2516 www promega com Printed in USA Part TM396 Revised 1 20 Page 19 2013 2020 Promega Corporation All Rights Reserved QuantiFluor is a registered trademark o...

Страница 21: ...Promega Corporation Promega Corporation 2800 2800 W Woods Hollow Road oods Hollow Road Madison WI 53711 5399 USA Phone 608 274 4330 TM396 0120_EIVD_TM qxd 1 27 2020 2 50 PM Page b...

Отзывы: