background image

5.

Troubleshooting (continued) 

Symptoms

Causes and Comments

 

Quantus™ Fluorometer powered up, 

Unplug USB Cord from the instrument, and 

but the screen was blank or could not 

plug the cord back in, making sure no buttons  

be deciphered or did not respond 

are pressed during this time. 

when buttons are pressed 

Quantus™ Fluorometer was plugged in,  Instrument entered sleep mode.  Press any 
but the screen was dark 

button to wake up the instrument. 

5.B. Error Messages 

Codes or Messages

Causes and Comments

 

Incomplete (on Calibration screen)

The protocol was not calibrated. Calibrate the  
protocol from the Calibration screen. See  
Section 3.A or 3.C. 

Invalid (on Calibration screen)

The expected standard-to-blank ratio was not  
met. Check the preparation of standard and  
blank samples to ensure that they were  
prepared correctly. Prepare new blank and  
standard samples if necessary. Measure  
fluorescence again. Press “Save” after  
measuring the fluorescence of the new standard  
and blank samples to retain the calibration  
values. 

Valid (on Calibration screen)

The standard-to-blank ratio was within the  
parameters expected by the instrument. Press  
“Save” to retain the calibration values, and  
proceed to measure fluorescence of the  
unknown samples. 

Lower than Blank (on the Main screen) The unknown sample was lower than that of the  

blank sample. LOW will appear on the History  
screen, but the raw RFU values will still be  
displayed. This indicates no sample was present  
in the tube holder or there was a problem with  
sample preparation. 

Saturated (on the Main screen)

Fluorescence was beyond the linear range of the  
instrument. SAT will appear on the History  
screen for both the concentration and raw RFU  
value. Dilute the sample or standard, and  
measure fluorescence again. 

Higher than Standard

The unknown sample was over 10% higher than  
that of the standard sample. The raw RFU values 
 will still be displayed on the history screen. 

Promega Corporation

 

·

 2800 Woods Hollow Road 

·

 Madison, WI 53711-5399 USA  

Toll Free in USA 800-356-9526 

·

 Phone 608-274-4330 

·

 Fax 608-277-2516 

·

 

www.promega.com

 

Part# TM396

Printed in USA. 

Page 16

Revised 1/20

TM396.0120_EIVD_TM.qxd  1/27/2020  2:50 PM  Page 16

Содержание E6150

Страница 1: ...T e c h n i c a l M a n u a l Quantus Fluorometer Operating Manual INSTRUCTIONS FOR USE OF PRODUCT E6150 PRINTED IN USA Revised 1 20 Part TM396 TM396 0120_EIVD_TM qxd 1 27 2020 2 50 PM Page a...

Страница 2: ...shooting 15 A Common Problems 15 B Error Messages 16 6 Appendix 17 A Updating Firmware 17 B Warranty and Service 17 C Certificate of Decontamination 18 D Related Products 19 7 Summary of Change 19 1 I...

Страница 3: ...y Includes Quantus Fluorometer USB Power Supply USB Cord Outlet Adapter Kit 4 adapters supplied for international compatibility Quick Reference Guide Operating Manual Product Registration Card Shippin...

Страница 4: ...on sensor Excitation Peak Red 625nm Blue 470nm Excitation Filters Red 640nm shortpass wavelengths up to 640nm Blue 495nm shortpass wavelengths up to 495nm Emission Filters Red 660 720nm Blue 510 580nm...

Страница 5: ...ques Veuillez contacter votre repr sentant Promega local concernant la destruction de l appareil Veuillez respecter les exigences de votre tablissement concernant la destruction des accessoires 9630TA...

Страница 6: ...nt may be impaired Do not disassemble the unit The Quantus Fluorometer accepts 0 5ml PCR tubes Do not force larger tubes into the tube holder Use caution around solvents because they may corrode the p...

Страница 7: ...etting Up the Quantus Fluorometer 1 Position the Quantus Fluorometer on a flat surface 2 Select the appropriate outlet adapter for your wall outlet from the Outlet Adapter Kit provided Promega Corpora...

Страница 8: ...un setup exe Follow prompts to install the software If NET 4 0 framework is not installed on the computer the software installer will prompt the user to download and install the framework software ava...

Страница 9: ...s Fluorometer protocol has calibration values associated with the specific assay reagent Once the protocol of interest is calibrated the Quantus Fluorometer will save the calibration values for future...

Страница 10: ...0 single concentration standard curve 1 400 RNA Cat E3310 high concentration standard curve 1 400 RNA Cat E3310 low concentration standard curve 1 2 000 ssDNA Cat E3190 high concentration standard cur...

Страница 11: ...electing the desired protocol navigate to the Calibration screen 2 Place the blank sample into the tube holder and close the lid Select Read Blank and the fluorescence in relative fluorescence units R...

Страница 12: ...tered is 100 200 0 5 4 Choose the appropriate fluorescence channel Red Excitation 640nm shortpass emission 660 720nm Blue Excitation 495nm shortpass emission 510 580nm 5 Select Save Note To edit or de...

Страница 13: ...it is within the linear range of the standard 1 Press any button to wake up the Quantus Fluorometer The Home screen will appear displaying the last protocol used the last concentration measurement an...

Страница 14: ...esult history select Clear from the menu list Results can be transferred to a connected computer as the measurements are read using the Quantus Software see Section 2 B The default fields transferred...

Страница 15: ...f additional cleaning is needed wipe with a cloth dampened with a mild detergent remove detergent with a cloth dampened with water and wipe dry 4 Do not use organic solvents they may damage the plasti...

Страница 16: ...ow concentration then use the low concentration standard and that the standard used is similar to the sample i e if the unknown sample is plasmid DNA then use a plasmid DNA standard Sample concentrati...

Страница 17: ...o retain the calibration values Valid on Calibration screen The standard to blank ratio was within the parameters expected by the instrument Press Save to retain the calibration values and proceed to...

Страница 18: ...e update only when instructed by Promega Technical Services Follow the instructions located at www promega com quantusfluorometerfirmware 6 B Warranty and Service The Quantus Fluorometer comes with a...

Страница 19: ...aminated and can be handled without exposing personnel to health hazards Circle the type of material used in the instrument Chemical Biological Radioactive Briefly describe the decontamination procedu...

Страница 20: ...00 356 9526 Phone 608 274 4330 Fax 608 277 2516 www promega com Printed in USA Part TM396 Revised 1 20 Page 19 2013 2020 Promega Corporation All Rights Reserved QuantiFluor is a registered trademark o...

Страница 21: ...Promega Corporation Promega Corporation 2800 2800 W Woods Hollow Road oods Hollow Road Madison WI 53711 5399 USA Phone 608 274 4330 TM396 0120_EIVD_TM qxd 1 27 2020 2 50 PM Page b...

Отзывы: