AUTOSETUP
If AUTOSETUP is necessary, it shall only be executed under the Manufacturer’s
authorization.
Following procedure sets all programmable values which are same as the ones indicated in
the column of relevant model in programming table.
To do that, it is necessary:
•
to disconnect the machine from power supply;
•
to hold Decrease button pressed
and power the machine again;
•
to keep on holding the Decrease button pressed. The display shows:
AUTOSETUP
and then:
Model 03
EASYMIX 120 PLUS
•
At this point release the key.
•
The machine will automatically set to Stop.
WARNING: With every autosetup, probe offsets are reset. Therefore, we recommend
checking the programming table before executing any autosetup and then to set all
offset values again.