Rescue Life User Manual vers. 3.0 Feb. 2015
62
230J,175ohm
For impedance value of 40ohm or less the maximum energy is 150J. If the energy is set to a
value higher then 150J the machine will set automatically the energy to 150J.
For impedance values different from 50 ohm the accuracy of the released energy is +/- 15%;
For impedance of 50 ohm the accuracy is +/-
10%.
IMPENDANCE LIMITS
RESCUE LIFE does not release the shock if the patient impedance is less than 20 ohm or over
200 ohm.
SYNC/NO SYNC MODE
When the
RESCUE LIFE
is on it is automatically set on no-sync mode.
Only the operator can set the sync/no-sync mode which is clearly displayed on the screen. The
device can not automatically set the sync mode.
In the sync mode the device releases the defibrillation shock only when the ‘R’ peak in the
‘QRS’ complex is detected.
The maximum response time between the “R” peak and the defibrillation shock is less than
60ms.
CHARGING TIME TO ACHIEVE THE MAXIMUM ENERGY (230JOULE)
When the device is connected to the AC supply (nominal AC voltage) and batteries are 100%
charged :< 7 sec.
When the device is connected to the AC supply (AC voltage 90%) after 15 shocks :< 10 sec.
Содержание Rescue Life
Страница 1: ...RESCUE LIFE EXTERNAL BIPHASIC DEFIBRILLATOR AND MONITOR USER MANUAL V 3 0 Feb 2015...
Страница 2: ...Rescue Life User Manual vers 3 0 Feb 2015...
Страница 6: ...Rescue Life User Manual vers 3 0 Feb 2015...
Страница 31: ...Rescue Life User Manual vers 3 0 Feb 2015 25 Precordial lead electrodes sites for the 10 wires ECG cable...
Страница 63: ...Rescue Life User Manual vers 3 0 Feb 2015 57...
Страница 74: ......
Страница 75: ......
Страница 76: ...Rescue Life User Manual vers 3 0 Feb 2015 www progettimedical com...