Rescue Life User Manual vers. 3.0 Feb. 2015
24
HOW TO USE THE ECG PATIENT CABLE CONNECTION AND ELECTRODES
PLACEMENT
The ECG (electrocardiogram) is a recording of the electrical activity of the heart. The ECG is
obtained by placing either electrodes or paddles on the patient and allows the heart’s electrical
activity to be monitored and recorded. ECG monitoring allows to identify and interpret the
cardiac rhythms or dysrhythmias and calculation of heart rate.
ECG Electrode Requirements
Electrode quality is critical for obtaining an undistorted ECG signal. Always check the date code
on.
electrode packages for the Use By date before applying the electrodes to a patient. Do not use
electrodes with expired Use By date codes. Disposable electrodes are intended for a single use.
For best ECG monitoring results, use silver/silver chloride (Ag/AgCl) electrodes.
Other types of electrodes will display the post-defibrillation ECG in less time than expected.
Possible misinterpretation of ECG data.
The frequency response of the monitor screen is intended only for basic ECG rhythm
identification; it does not provide the resolution required for diagnostic and ST segment
interpretation. For diagnostic or ST segment interpretation, or to enhance internal pacemaker
pulse visibility, attach the ECG cable and then print the ECG rhythm.
ECG CONNECTION
Connect the patient cable in its proper port
‘ECG’,
placed on the frontal panel of the device.
T
HE
STANDARD
VERSION OF
RESCUE LIFE
HAS ONLY A
5
WIRE PATIENT CABLE
. W
HEN THE CABLE IS
CONNECTED TO THE MACHINE
,
IT WILL SWITCH AUTOMATICALLY TO THE
ECG
CABLE MONITORING
.
In the optional ECG 10 wires patient cable version the monitoring source is controlled only by
the selection (TRACE) on the operational screen. The 3wires, 5 wires and 10 wires ECG cable
can be used.
Follow the drawings below for the electrodes connection.
Limb lead electrodes placement:
Содержание Rescue Life
Страница 1: ...RESCUE LIFE EXTERNAL BIPHASIC DEFIBRILLATOR AND MONITOR USER MANUAL V 3 0 Feb 2015...
Страница 2: ...Rescue Life User Manual vers 3 0 Feb 2015...
Страница 6: ...Rescue Life User Manual vers 3 0 Feb 2015...
Страница 31: ...Rescue Life User Manual vers 3 0 Feb 2015 25 Precordial lead electrodes sites for the 10 wires ECG cable...
Страница 63: ...Rescue Life User Manual vers 3 0 Feb 2015 57...
Страница 74: ......
Страница 75: ......
Страница 76: ...Rescue Life User Manual vers 3 0 Feb 2015 www progettimedical com...