DEFIBRILLATION
DEFIBRILLATION PROCEDURE IN MANUAL
OR ADVISORY (ADV) MODE
1. Switch on the device pressing the
ON/OFF
button. Connect and lock the pads plug to
start the operation. The ECG signal will be displayed and by default the energy is set to
150 J.
2. Place the pads on the patient chest and analyze the ECG trace in order to decide if
defibrillation is necessary. If the advisory (ADV) mode is on, the device automatically
analyzes the ECG rhythm and it will advise you on the display if the defibrillation is
recommended or not.
3. Select the energy level required using the Speed Dial.
4. If you are using the standard pads press both push buttons on the handles to start
charging.
5. If you are using disposable pads press the
CHARGE
button on the device panel to start
charging.
6. On the screen the charge status bar indicates that the charging procedure is on; at the
same time the ascending sound will start.
7. When the charge ends the red light on the
SHOCK
button will turn on indicating that
RESCUE LIFE
is ready for defibrillation.
8. To release the defibrillation shock, press both push buttons on the standard
pads. If using disposable pads press the SHOCK key in the front panel to
release the defibrillation shock.
9. The shock has to be released within 30 sec from the charge completed; after 30 sec the
RESCUE LIFE will discharge internally.
10.If the defibrillation is not required, press the
DISARM (F1)
key to discharge internally.
11.If the SYNC function is on, SYNCHRONIZED CARDIOVERSION can be done.
W
HEN THE MESSAGE
ATTACH PADS
IS DISPLAYED ON THE SCREEN AND THE CHARGE IS STARTED
,
THE
DEVICE ASSUMES A STANDARD IMPEDANCE OF
50
OHM
. I
F THE MESSAGE
ATTACH PADS
PERSISTS
WHEN THE SHOCK HAS TO BE DELIVERED
,
THE DEVICE WILL DISCHARGE INTERNALLY
T
HE MESSAGE
ATTACH PADS
CAN BE DISPLAYED ALSO IN PRESENCE OF A NO SUFFICIENT ELECTRIC
CONTACT BETWEEN THE PADS AND THE PATIENT SKIN
;
IN THIS CASE ADD CONDUCTIVE GEL AND PRESS
STRONGLY THE DEFIBRILLATION PADS ON THE PATIENT SKIN
. D
URING
SYNC
MODE
,
THE SHOCK WILL
NOT BE RELEASED IF THE
ECG
TRACE IS NOT STABLE AND THE
QRS
COMPLEX IS NOT VALID
.
Содержание Rescue Life
Страница 1: ...RESCUE LIFE EXTERNAL BIPHASIC DEFIBRILLATOR AND MONITOR USER MANUAL V 3 0 Feb 2015...
Страница 2: ...Rescue Life User Manual vers 3 0 Feb 2015...
Страница 6: ...Rescue Life User Manual vers 3 0 Feb 2015...
Страница 31: ...Rescue Life User Manual vers 3 0 Feb 2015 25 Precordial lead electrodes sites for the 10 wires ECG cable...
Страница 63: ...Rescue Life User Manual vers 3 0 Feb 2015 57...
Страница 74: ......
Страница 75: ......
Страница 76: ...Rescue Life User Manual vers 3 0 Feb 2015 www progettimedical com...