Specifications JB-70
18
Specifications
JB-70 Dental X-Ray System
The following specifications contain information required to be provided to the user per Federal
Regulation 21 CFR.
Line Voltage
120 VAC +/- 10% 50/60 Hz
Max line current is 10 Amps
230 VAC +/- 10% 50/60 Hz
Max line current is 5 Amps
Load Regulation Must not exceed 6%
Tube Potential
70 KVp +/-10%
Tube Current
8 mA +/- 2 mA
Exposure Time
Selections
50 ms through 1.65 seconds for 60 Hz operation
60 ms through 2.0 seconds for 50 Hz operation.
Timer Accuracy
+/- 34 ms
Source to Skin
Distance
8 inch (20 cm)
12 inch (30 cm)
Radiation Output
Approximately 800 mR at the following sets
70 kVp 8 mA 1.0 sec 20 cm SID
Minimum Inherent
Filtration
2 mm Aluminum equivalent
Minimum Half Value
Layer
1.7 mm Aluminum equivalent at 70 kVp
Focal Spot
0.7 mm (IEC 336)
Duty Factor
1:60
Leakage Technique
Factors
0.25 mA at 70 kVp
Target Angle
20 degrees
Operating
Temperature
+50 F/+113 F (+10 C/+45 C)
Storage
Temperature
-31F/+150 F (-35 C/+66 C)
Maximum Altitude
12,000 ft
Cone Focal Length
8 inch(20 cm) for bisecting angle technique
12 inch(30 cm) for paralleling technique
Diameter of X-Ray
2.7 inch (6.9 cm) at the end of the Cone