Recommended Maintenance JB-70
12
Recommended Maintenance
Checking System Functions
Important Notice
The following checks must be performed at the installation of the JB-70 Dental X-Ray System and as part
of the recommended maintenance at least every twelve months after installation. Failure to perform these
checks may result in an installation that does not comply with U.S. Radiation Performance Standards 21
CFR Subchapter J.
If the JB-70 Dental X-Ray System does not perform the functions below, see the Troubleshooting
section of this manual or contact Progeny Dental
’s Technical Support. Do not use the system until
the problems are resolved.
Regular Maintenance
In the interest of equipment safety, a regular maintenance program must be
established. This maintenance program should consist of cleaning and
disinfecting as well as annual system function checking. It is the owner’s
responsibility to arrange for this service and to assure that the personnel
performing this are fully qualified to service Progeny Dental x-ray equipment.
Cleaning and Disinfecting
All external and cleanable surfaces can be cleaned and/or disinfected using the
following list of commonly used solutions:
•
Isopropyl alcohol
•
Phenylphenl
•
Benzyl-P-Chlorophenol
•
Iodophor
Note:
Use bleach or glutaraldehydes only occasionally.
Verify the contents of any cleaning or disinfecting solution against the list above
before applying it to any surface of the JB-70 Dental X-Ray System.