52 PRODIGIT
The starting point of the graphic data is random, not starting from 0 degrees
I GRAPH?
(For RS232 or USB Interface only)
Syntax: MEAS: IGRAPH?{; | NL}
Description: 3bytes x4096(Current Graph) +0x0D 0x0A
The 1st highest bit is positive and negative flag, 0=positive value 1=negative value, the
value contains decimal place, and the decimal place is determined according to the
current gear position (please refer to 1.6 Specifications)
Current Graph(IRange=10A):0x80 0x1F 0x40= -8.000A
※
Before reading , you need to send command
〝
LOCK ON
〞
to lock the data update.
At this time, the measurement data will not be updated. After reading , you need to send
command
〝
LOCK OFF
〞
to unlock the data.
The starting point of the graphic data is random, not starting from 0 degrees
W GRAPH?
(For RS232 or USB Interface only)
Syntax:MEAS:WGRAPH?{; | NL}
Description: 5bytes x4096(Watt Graph)+0x0D 0x0A
The watt decimal place is composed of voltage and current gear Decide
Watt Graph(400V 10A
檔位
):0x80 0x05 0x3E 0xC6 0x00=-880.00000W
※
Before reading , you need to send command
〝
LOCK ON
〞
to lock the data update.
At this time, the measurement data will not be updated. After reading , you need to send
command
〝
LOCK OFF
〞
to unlock the data.
The starting point of the graphic data is random, not starting from 0 degrees
AH?
Syntax:
MEAS: AH? {; | NL}
Description: Read Cumulative amp hour
PAV?
Syntax:
MEAS: PAV? {; | NL}
Description: Read Average Watt
AAV?
Syntax:
MEAS: AAV? {; | NL}
Description: Read Average Amp