44 PRODIGIT
Chapter 4 Communication Interface programming operation
4-1. Introduction
The rear panel Communication Interface programming of 4016 Series is designed to connect
PC or NOTEBOOK PC with Communication Interface programming; you can use an
application such as high-level syntax C and VB to form an automatic control system.
NOTE: When use USB interface controls the 4016 Series, the 4016 Series will convert
the USB interface to RS232 interface.
4-2. GPIB address modification instructions
4.2.1. Press the EDIT key under the Menu page to enter the Edit GPIB Settings page.
•
Press the Select key to enter the edit.
•
Number keys or up and down keys to set the value.
•
Press the Select key to finish editing.
•
Press the Menu key to leave the GPIB screen.
•
After modifying the GPIB address and returning to the MENU page, you need to
power off the 4016 and reboot to update the GPIB address.
4-3. RS232 protocol
The following RS232 commands are same as GPIB commands. The RS232 protocol in 4016
Series mainframe is listing below:
Baud-rate
:
115200bps
Parity
:
None
Data bit
:
8 bits
Stop bit
:
1 bit
Handshaking
:
Hardware (RTS/CTS).
The RS232 Interface connector of 4016 Series rear panel, RS232 is shown in Fig4-1.