4016 Series Operation Manual 7
2-5 Environmental requirements
•
Indoor use.
•
Measurement Category I.
•
Pollution Degree 2.
•
Relative Humidity 80% Max.
•
Ambient Temperature 0 to +40°C
•
Altitude up to 2000m.
•
The equipment is not for measurements performed for CAT II, III and IV.
•
Transient Overvoltage on the mains supply can be 2500V.
2-6 Repair
If the instrument is damaged, please attach a tag to the instrument to identify the owner and
indicated the require service or repairing. And inform the Prodigit sales and service office or
representative.
Fuses (F1, F2, Fuse 25A), overcurrent protection diodes (D1~4, D31~32) and switching
components (14016402) are consumables and are not covered by the warranty.
2-7 Cleaning
2-8 Power Up
The following procedure should be followed before applying mains power:
2.8.1. Turn off (O) the POWER switch
2.8.2. Check that the power cord is corrected.
2.8.3. Check that nothing is connected to the DC INPUT on the rear panels.
2.8.4. Turn on POWER switch.
Use a soft or slightly damp cloth to clean this product.
BEFORE you clean the unit, switch the mains power off and disconnect the input lead.
・
Please do NOT use any organic solvent capable of changing the nature of the plastic
such as benzene or acetone.
・
Please ensure that no liquid is allowed to penetrate this product.