48 PRODIGIT
VAL
Purpose
:
:
:
:
To set the lower limit value of threshold power (VA).
Command syntax
:
:
:
:
VAL{SP}{NR2}{
;
|NL}
Description
:
:
:
:
This command is to set the lower limit value of threshold power (VA). When power
(VA) is lower than this lower limit value, NG indicating light will come on to indicate
〝
NO GOOD
〞
.
For Example
:
:
:
:
VAL 0.05 ; To set the lower limit value of threshold power (VA) is 0.05 VA.
VAH
Purpose
:
:
:
:
To set the upper limit value of threshold power (VA).
Command syntax
:
:
:
:
VAH{SP}{NR2}{
;
|NL}
Description
:
:
:
:
This command is to set the upper limit value of threshold power (VA). When power
(VA) is higher than this upper limit value of threshold power (VA), NG indicating light
will come on to indicate
〝
NO GOOD
〞
.
For Example
:
:
:
:
VAH 250.0 ; To set the upper limit value of threshold power (VA) is 250.0 VA.
AFREQ
Purpose
:
:
:
:
To set auto frequency detect function.
Command syntax
:
:
:
:
AFREQ
{SP}{0|1|OFF|ON}{
;
|NL}
Description:
1.Enable auto frequency detect function (AFREQ ON)
:
3260A series electronic load
will auto frequency external signal to control load current and voltage synchronous.
At this time if place “FREQ” command, it will be ineffective.
2.Disable auto frequency detect function (AFREQ OFF)
:
at this time user needs to
make frequency setting. When AFREQ OFF, the frequency will be fixed on the last
detective frequency.
For Example
:
:
:
:
a.AFREQ ON
b.AFREQ OFF
REMOTE
Purpose
:
:
:
:
Go to remote.
Command syntax
:
:
:
:
REMOTE{
;
|NL}
Description
:
:
:
:
Let control procedure enter into remote state.